Yang G Z, Burger P, Kilner P J, Karwatowski S P, Firmin D N
Magnetic Resonance Unit, Royal Brompton Hospital, London, United Kingdom.
J Magn Reson Imaging. 1996 May-Jun;6(3):495-502. doi: 10.1002/jmri.1880060313.
A motion-registered spatiotemporal phase-unwrapping method for extending the dynamic range of cine magnetic resonance phase velocity measurements is presented. The interframe cardiac movement is estimated from the magnitude image derived from the velocity encoded raw data, which ensures that the phase signal is unwrapped in the temporal direction with reference to pixels belonging to the same anatomic flow region. An extra step of spatial phase correction is then used to further eliminate any residual errors. The combination of spatial and temporal information for phase unwrapping provides a robust way of extending the dynamic range of cine velocity data in the presence of large phase aliasing errors.
提出了一种用于扩展电影磁共振相速度测量动态范围的运动注册时空相位解缠方法。从速度编码原始数据导出的幅度图像估计帧间心脏运动,这确保了相位信号在时间方向上相对于属于相同解剖流动区域的像素进行解缠。然后使用额外的空间相位校正步骤来进一步消除任何残留误差。在存在大相位混叠误差的情况下,用于相位解缠的空间和时间信息的组合提供了一种扩展电影速度数据动态范围的稳健方法。