Signal profile measurements for evaluation of the volume-selection performance of ISIS.

作者信息

Ljungberg M, Starck G, Forssell-Aronsson E, Alpsten M, Ekholm S

机构信息

Department of Radiation Physics, Göteborg University, Sahlgrenska University Hospital, Sweden.

出版信息

NMR Biomed. 1995 Sep;8(6):271-7. doi: 10.1002/nbm.1940080607.

Abstract

High-resolution signal profiles obtained with a test phantom were used in this study to evaluate the volume-selection performance of an implementation of ISIS (Image Selected In vivo Spectroscopy). The phantom simulated the brain with regard to volume and loading of coil. A remotely controlled, movable signal source inside the phantom was filled with orthophosphoric acid. Signal profiles of the volume of interest (VOI) were measured in three perpendicular directions. Special interest was focused on the transition zones, the position of the profiles, and the effects of off-resonance and T1 smearing. The transition zones were on average 5.6 mm wide and the full width at half maximum (FWHM) was 35 mm for a VOI of 40 x 40 x 40 mm3. The positions of the centre of the signal profiles were x = 3.2, y = -0.7 and z = 3.3 mm off-centre. The deviation of the volume position could be explained by off-resonance effects during imaging and spectroscopy. These data illustrate the importance of detailed knowledge of the volume-selection performance when attempting precision measurements using image-guided in vivo MRS.

摘要

本研究使用通过测试体模获得的高分辨率信号剖面图来评估ISIS(体内光谱成像选择)实施方法的体积选择性能。该体模在体积和线圈负载方面模拟大脑。体模内部的一个遥控可移动信号源填充有正磷酸。在三个垂直方向上测量感兴趣体积(VOI)的信号剖面图。特别关注过渡区、剖面图的位置以及失谐和T1模糊效应。对于40×40×40立方毫米的VOI,过渡区平均宽度为5.6毫米,半高宽(FWHM)为35毫米。信号剖面图中心位置偏离中心的距离为x = 3.2毫米、y = -0.7毫米和z = 3.3毫米。体积位置的偏差可以通过成像和光谱过程中的失谐效应来解释。这些数据说明了在尝试使用图像引导的体内磁共振波谱进行精确测量时,详细了解体积选择性能的重要性。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验