Rohlmann A, Bergmann G, Graichen F, Mayer H M
Department of Biomechanics, Free University Berlin, Germany.
Spine (Phila Pa 1976). 1995 Dec 15;20(24):2683-9. doi: 10.1097/00007632-199512150-00009.
In the present study, the loads in an internal spinal fixation device were measured in vivo.
To determine the implant loads for different activities before and after additional anterior stabilization of the spine.
Mathematical models exist for predicting spinal loads. The intradiscal pressure has been measured for many body positions and activities. The loads on internal spinal fixation devices have not been measured before in vivo.
Telemeterized AO spinal internal fixators were implanted in a patient with degenerative instability. The implants allow the in vivo measurement of three force components and three moments acting in the implant.
When the patient was lying in relaxed positions, the implant loads were small. Before additional anterior stabilization, the loads were also small for sitting, standing, and walking. The bending moment in the sagittal plane was less than 3 Nm for these activities. The highest loads within the first 4 weeks after implantation were measured while the patient turned from a supine to a lateral position against the advice of the physiotherapist. After anterior stabilization, the maximum loads for the relaxed lying positions were altered only slightly. Much higher axial forces and bending moments were measured for sitting, standing, and walking. The maximum bending moment increased to 5-8 Nm for these activities. The implant loads for sitting were not higher than for standing.
Flexion and lateral bending of the upper body and weight-carrying during sitting, standing, or walking should be avoided in the first few months after anterior stabilization.
在本研究中,对体内的脊柱内固定装置的负荷进行了测量。
确定脊柱附加前路稳定术前后不同活动时的植入物负荷。
存在用于预测脊柱负荷的数学模型。已对许多身体姿势和活动时的椎间盘内压力进行了测量。此前尚未在体内测量过脊柱内固定装置上的负荷。
将遥测式AO脊柱内固定器植入一名患有退行性不稳定的患者体内。这些植入物能够在体内测量作用于植入物的三个力分量和三个力矩。
当患者处于放松体位时,植入物负荷较小。在附加前路稳定术之前,坐、站和行走时的负荷也较小。这些活动时矢状面内的弯矩小于3牛米。在植入后的前4周内,测得的最高负荷是在患者违背物理治疗师的建议从仰卧位转为侧卧位时。前路稳定术后,放松卧位时的最大负荷仅略有改变。坐、站和行走时测得的轴向力和弯矩要高得多。这些活动时的最大弯矩增加到5 - 8牛米。坐时的植入物负荷不高于站时。
在前路稳定术后的头几个月内,应避免上半身的屈曲和侧屈以及坐、站或行走时负重。