Bedside measurement of the third ventricle's diameter during episodes of arising intracranial pressure after head trauma. Using transcranial real-time sonography for a non-invasive examination of intracranial compensation mechanisms.

作者信息

Mursch K, Vogelsang J P, Zimmerer B, Ludwig H C, Behnke J, Markakis E

机构信息

Department of Neurosurgery, Georg-August-University, Götingen, Federal Republic of Germany.

出版信息

Acta Neurochir (Wien). 1995;137(1-2):19-23; discussion 23-4. doi: 10.1007/BF02188774.

Abstract

Using transcranial real-time sonography, changes in the axial diameter of the third ventricle during manoeuvres, which increased intracranial pressure (ICP), were measured in 28 patients with moderate to severe head injury. The measurements were correlated with ICP measured by epidural pressure monitoring. We observed reductions in diameter ranging from 0.3 to 1.1 mm with rises in intracranial pressure of at least 5 mm in 22 patients of whom only one died. In 6 additional patients, no changes in diameter were seen, and 5 of the 6 died. We interpret that poor outcome as a measurable inability for the brain to expel cerebrospinal fluid into extracerebral compartments during increased ICP. Transcranial real-time sonography may provide additional information about intracranial cerebral fluid dynamics and compliance.