Winrow D, Beckmann C F, Lacomis J M, Dedrick C G
Department of Diagnostic Radiology, Lahey Clinic, 41 Mall Road, Burlington, MA 01805, USA.
Cardiovasc Intervent Radiol. 1996 Jul-Aug;19(4):275-7. doi: 10.1007/BF02577649.
Two cases are presented in which a pigtail catheter was entrapped by the chordae tendineae of the tricuspid valve during pulmonary arteriography. A technique for removal of the catheter from its entanglement by the chordae tendineae is described. Caution must be taken when advancing through the right ventricle a catheter that appears to be entrapped by the chordae tendineae. When such an entanglement occurs, measures to reduce the risk of rupturing a papillary muscle must be taken.
本文报告两例在肺动脉造影期间猪尾导管被三尖瓣腱索卡压的病例。描述了一种将导管从腱索缠绕中取出的技术。当推进似乎被腱索卡压的导管通过右心室时必须小心。当发生这种缠绕时,必须采取措施降低乳头肌破裂的风险。