Cohen A E, Kerdahi K F
U.S. Food and Drug Administration, Northeast Regional Laboratory, Brooklyn, NY 11232, USA.
J AOAC Int. 1996 Jul-Aug;79(4):858-60.
The Vitek Immunodiagnostic Assay System (VI-DAS) for Escherichia coli O157, a rapid and fully automated test, was evaluated for use in detecting the foodborne pathogen E. coli O157:H7 in soft, semisoft, and hard cheeses. Sixty-five cheese samples were artificially contaminated at low (2-4 colonyforming units [cfu]/25 g) and high (7-10 cfu/25 g) levels with one of 2 strains of enterohemorrhagic E. coli O157:H7. Contamination at high levels was detected in all cheeses by VIDAS, whereas in 5 cheeses (7.7%) inoculated at low levels, contamination was not detected. In 15 additional cheeses inoculated with cold-stressed cells, both VIDAS and the Bacteriological Analytical Manual cultural assay detected all high and low levels of contamination. No false positives or interference from product background fluorescence was encountered in any of the cheeses tested by VIDAS.
用于检测大肠杆菌O157的Vitek免疫诊断分析系统(VI-DAS)是一种快速且全自动的检测方法,该方法被评估用于检测软质、半硬质和硬质奶酪中食源性病原体大肠杆菌O157:H7。65个奶酪样品分别被两种肠出血性大肠杆菌O157:H7菌株之一以低水平(2 - 4个菌落形成单位[cfu]/25克)和高水平(7 - 10 cfu/25克)进行人工污染。VIDAS检测到所有奶酪中的高水平污染,而在5个低水平接种的奶酪(7.7%)中未检测到污染。在另外15个接种了冷应激细胞的奶酪中,VIDAS和《细菌学分析手册》中的培养检测法均检测到了所有高水平和低水平的污染。在VIDAS检测的任何奶酪中均未遇到假阳性或产品背景荧光干扰。