Zhang C, Lin Q, Cao C
School of Stomatology, Beijing Medical University.
Zhonghua Kou Qiang Yi Xue Za Zhi. 1995 Nov;30(6):340-2, 384.
Two methods were tested for measuring the sensitivity of dentin to mechanical (force-sensitive probe) and cold air stimuli. Of the 212 sensitive teeth in 71 patients, 94.8% were sensitive to probe, with a mean threshold of 22.79g; 89.6% were sensitive to cold air. These two methods were significantly correlated (P < 0.001). The occlusal sites were less sensitive than the buccal sites measured by probe and cold air. It suggests that force-sensitive probe is a simple, quantitative method of measuring tooth hypersensitivity.
测试了两种测量牙本质对机械刺激(力敏探针)和冷空气刺激敏感度的方法。在71名患者的212颗敏感牙中,94.8%的牙齿对探针敏感,平均阈值为22.79克;89.6%的牙齿对冷空气敏感。这两种方法显著相关(P<0.001)。通过探针和冷空气测量发现,咬合部位的敏感度低于颊侧部位。这表明力敏探针是一种测量牙齿过敏症的简单定量方法。