Variability in estimation of coronary dimensions from 6F and 8F catheters.

作者信息

Legrand V, Raskinet B, Martinez C, Kulbertus H

机构信息

Division of Interventional Cardiology, C.H.U. Sart-Tilman, Liege, Belgium.

出版信息

Cathet Cardiovasc Diagn. 1996 Jan;37(1):39-45; discussion 46. doi: 10.1002/(SICI)1097-0304(199601)37:1<39::AID-CCD9>3.0.CO;2-5.

Abstract

To investigate the suitability of diagnostic 6F catheters for coronary angiographic measures in the clinical setting, we determined the relative accuracy and reproducibility of the measures obtained with these catheters as scaling devices in 59 stenoses. Comparison was made with duplicate injections, obtained before angioplasty, using an 8F guiding catheter as scaling device. Intra- and interobserver variability was evaluated in 15 stenoses. The coefficient of variation averaged 18.3% for the minimal lumen diameter, 10.4% for the percent stenosis, and only 7.4% for the reference diameter. Reproducibility of angiographic measures done with the 6F catheter was similar to that obtained with the 8F catheter, although accuracy was lower with the 6F for the measurement of reference diameter. Thus, quantitative coronary angiography (QCA) measures derived from routine diagnostic angiograms may be suitable for determination of reference diameter, allowing enough precision for determination of the size of a coronary device for intervention, but these measures may lack accuracy for precise determination of minimum diameter and percent stenosis, making their use questionable in studies looking at individual changes in coronary stenosis dimensions.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验