Accuracy and reproducibility of quantitative coronary arteriography using 6 and 8 French catheters with cine angiographic acquisition.

作者信息

Ellis S G, Pinto I M, McGillem M J, DeBoe S F, LeFree M T, Mancini G B

机构信息

Department of Internal Medicine, University of Michigan Hospital, Ann Arbor.

出版信息

Cathet Cardiovasc Diagn. 1991 Jan;22(1):52-5. doi: 10.1002/ccd.1810220113.

Abstract

To determine the suitability of 6 French catheters for quantitative coronary arteriography, the relative accuracy and reproducibility of one type of these catheters was compared to that obtained with standard 8 French catheters in 20 stenoses. Duplicate injections with polyurethane 6 French catheters were obtained using hand and power injection technique with cineangiographic acquisition (four 6 French catheter injections total per stenosis). Measurements of both percent diameter stenosis and absolute dimensions were compared to those obtained with hand injection and cine acquisition using 8 French catheters as a "gold standard." While the reproducibility of dimension determination with the 6 French catheter was generally similar to that obtained with the 8 French catheter (0.27 +/- 0.23 mm for absolute diameter and 8.1 +/- 7.4% for percent diameter stenosis), accuracy was significantly less for the 6 French catheter for measurement of absolute dimensions. Thus, while apparently well suited for serial measurements of the same stenoses, 6 French catheters may not be as accurate in the determination of absolute artery dimensions as 8 French catheters.