Half-cone beam collimation for triple-camera SPECT systems.

作者信息

Li J, Jaszczak R J, Van Mullekom A, Scarfone C, Greer K L, Coleman R E

机构信息

Department of Radiology, Duke University Medical Center, Durham, North Carolina 27710, USA.

出版信息

J Nucl Med. 1996 Mar;37(3):498-502.

PMID:8772655
Abstract

UNLABELLED

Cone-beam collimators provide increased sensitivity at similar resolution compared to other collimators. The use of cone-beam collimators for brain imaging with triple-camera SPECT systems, however, results in truncation of the base of the brain because of clearance of the shoulders. A half-cone beam collimator does not have the problem of truncation. The objective of this study was to compare the performance characteristics of half-cone beam with parallel-beam and fan-beam collimators with similar resolution characteristics for SPECT imaging of the brain.

METHODS

A half-cone beam collimator with the focal point located towards the base of the brain was built for a triple-camera SPECT system. Spatial resolutions and sensitivities of three collimators were measured.

RESULTS

When 10-cm from the collimator surface, the planar spatial resolutions FWHM in mm (point source sensitivities in cps-MBq) for half-cone beam, fan-beam and parallel-beam collimators were 5.2 (85.6), 5.1 (55.6) and 5.9 (39.7), respectively. Image quality was evaluated using a three-dimensional Hoffman brain phantom and patient data. The deeper gray matter were more clearly visualized in the half-cone beam scans.

CONCLUSION

Half-cone beam collimation provides higher sensitivity and offers the potential for improved brain imaging compared with parallel-beam and fan-beam collimation when used with a triple-camera SPECT system.

摘要

未标注

与其他准直器相比,锥束准直器在相似分辨率下可提供更高的灵敏度。然而,在使用三相机单光子发射计算机断层扫描(SPECT)系统进行脑成像时,由于肩部间隙,锥束准直器会导致脑底部截断。半锥束准直器不存在截断问题。本研究的目的是比较半锥束准直器与具有相似分辨率特性的平行束和扇形束准直器在脑SPECT成像中的性能特征。

方法

为三相机SPECT系统构建了一个焦点朝向脑底部的半锥束准直器。测量了三种准直器的空间分辨率和灵敏度。

结果

在距准直器表面10厘米处,半锥束、扇形束和平行束准直器的平面空间分辨率半高宽(mm)(点源灵敏度,cps/MBq)分别为5.2(85.6)、5.1(55.6)和5.9(39.7)。使用三维霍夫曼脑模型和患者数据评估图像质量。在半锥束扫描中,深部灰质更清晰可见。

结论

与平行束和扇形束准直相比,当与三相机SPECT系统一起使用时,半锥束准直可提供更高的灵敏度,并具有改善脑成像的潜力。