外周球囊血管成形术前后血管内超声图像定性和定量分析的观察者间可重复性

Interobserver reproducibility of qualitative and quantitative analysis of intravascular ultrasound images before and after peripheral balloon angioplasty.

作者信息

van der Lugt A, Gussenhoven E J, Pasterkamp G, Bom N, Posthuma D J, Stijnen T

机构信息

University Hospital Rotterdam-Dijkzigt, The Netherlands.

出版信息

Ultrasound Med Biol. 1996;22(4):399-404. doi: 10.1016/0301-5629(96)00034-8.

Abstract

In this study, interobserver agreement on intravascular ultrasound data obtained before and after balloon angioplasty (PTA) of the superficial femoral artery was assessed. Two observers analyzed intravascular ultrasound cross-sections from 38 patients. Interobserver agreement was good for soft and hard lesions, dissection and vascular damage (kappa values 0.61, 0.67, 0.69 and 0.66, respectively); moderate for eccentric lesions (kappa 0.45); fair for media ruptures (kappa 0.25); and poor for lipid deposits and plaque ruptures (kappa 0.0 and 0.04, respectively). Differences for the arc of normal wall and hard lesion were not significant, but were for dissection. There were no significant interobserver differences between area measurements. The coefficient of variation for free lumen area and media-bounded area before PTA was 17.2% and 10.5% and after PTA 11.2% and 9.2%, respectively. This study identified the intravascular ultrasound parameters that are reproducible.

摘要

在本研究中,评估了股浅动脉球囊血管成形术(PTA)前后获得的血管内超声数据的观察者间一致性。两名观察者分析了38例患者的血管内超声横截面。观察者间一致性在软斑块和硬斑块、夹层和血管损伤方面良好(kappa值分别为0.61、0.67、0.69和0.66);在偏心性病变方面为中等(kappa值为0.45);在中膜破裂方面为一般(kappa值为0.25);在脂质沉积和斑块破裂方面较差(kappa值分别为0.0和0.04)。正常管壁弧度和硬斑块的差异不显著,但夹层的差异显著。面积测量方面观察者间无显著差异。PTA前自由管腔面积和中膜包绕面积的变异系数分别为17.2%和10.5%,PTA后分别为11.2%和9.2%。本研究确定了可重复的血管内超声参数。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验