Suppr超能文献

血管内支架锚定主动脉移植物:小型猪体内自膨胀支架与球囊扩张支架的比较

Endovascular stent-anchored aortic grafts: a comparison between self-expanding and balloon-expandable stents in minipigs.

作者信息

Lindh M, Malina M, Ivancev K, Brunkwall J, Lindblad B

机构信息

Department of Diagnostic Radiology, Malmö University Hospital, Lund University, Sweden.

出版信息

J Endovasc Surg. 1996 Aug;3(3):284-9. doi: 10.1177/152660289600300307.

Abstract

PURPOSE

To study endovascular graft attachment with self-expanding Gianturco Z-stents and balloon-expanded Palmaz stents and the effect of these devices on the renal ostia.

METHODS

Ten stent-grafts were constructed, 5 with Gianturco Z-stents and 5 with Palmaz stents. The endografts were implanted under fluoroscopic guidance into the abdominal aorta of 10 pigs so that the uncovered portion of the proximal stent extended over the renal artery orifices. Distal aortic blood pressure and flow were measured before and after graft placement and 1 hour postprocedure. The aorta was then exposed surgically, and the central portion of the stent-graft was inspected through an aortotomy to assess perigraft leakage.

RESULTS

Stent-graft implantation was accurate and hemostatic in all cases, despite longitudinal folding of the graft due to oversizing. However, transverse folds produced pressure gradients (> 15 mmHg) between the ends of the graft in two cases. In another case, a pressure gradient resulted from partial thrombosis of the graft. In two cases, renal artery occlusion and thrombosis occurred due to coverage by the graft material. In two other animals, one of the renal arteries was entirely uncovered by a stent. The remaining 16 renal arteries were covered by the proximal stent but not the graft, as intended. One (6.25%) of these arteries thrombosed, but the remainder were grossly patent when the animals were sacrificed at 1 hour.

CONCLUSIONS

Both Palmaz and Gianturco Z-stents produced hemostatic endovascular graft attachment, even in the presence of moderate graft oversizing. The risk of acute renal artery occlusion from juxtarenal stenting does not appear to be prohibitive, but longer term observations are needed.

摘要

目的

研究自膨式Gianturco Z形支架和球囊扩张式Palmaz支架与血管内移植物的附着情况以及这些装置对肾动脉开口的影响。

方法

构建了10个支架移植物,其中5个使用Gianturco Z形支架,5个使用Palmaz支架。在荧光镜引导下将这些腔内移植物植入10头猪的腹主动脉,使近端支架的未覆盖部分延伸至肾动脉开口上方。在植入移植物前后及术后1小时测量远端主动脉血压和血流。然后通过手术暴露主动脉,通过主动脉切开术检查支架移植物的中心部分以评估移植物周围渗漏情况。

结果

在所有病例中,尽管由于移植物尺寸过大导致纵向折叠,但支架移植物植入准确且具有止血作用。然而,在两例中横向折叠在移植物两端之间产生了压力梯度(>15 mmHg)。在另一例中,压力梯度是由移植物部分血栓形成导致的。在两例中,由于移植物材料覆盖导致肾动脉闭塞和血栓形成。在另外两只动物中,其中一条肾动脉完全未被支架覆盖。其余16条肾动脉如预期那样被近端支架覆盖但未被移植物覆盖。其中一条(6.25%)动脉发生血栓形成,但在1小时后处死动物时其余动脉大体通畅。

结论

Palmaz支架和Gianturco Z形支架均能实现血管内移植物的止血附着,即使在移植物存在适度尺寸过大的情况下也是如此。肾旁支架置入导致急性肾动脉闭塞的风险似乎并非过高,但需要进行长期观察。