Bucx M J, van der Vegt M H, Snijders C J, Stijnen T, Wesselink P R
Department of Anaesthesia, University of Amsterdam, The Netherlands.
Can J Anaesth. 1996 Jul;43(7):665-71. doi: 10.1007/BF03017948.
In this study the effect of level of experience of the intubator on the forces applied by the Macintosh laryngoscope on the maxillary incisors in both the axial and transverse direction were investigated.
Five groups of different levels of experience (15 per group), staff anaesthetists, residents, nurse anaesthetists, surgeons and students, performed one laryngoscopy and subsequent intubation on an intubation manikin equipped with sensors to measure these forces.
Maximal transverse forces oriented toward the base of the skull (Fmtpmax) were between 0 and 10 N in 46 cases (61%), between 10 and 20 N in 21 (28%) and > or = 20 N (maximum 46.5 N) in eight cases (11%). The maximal values of the transverse forces oriented toward the intubator (Fmtnmax) were between 0 and 10 N in 69 cases (92%), between 10 and 20 N in 3 (4%) and > or = 20 N (maximum 36.4 N) in 3 (4%). Level of experience was related to Fmtpmax (Spearman: P = 0.04, r = 0.24). In addition, Fmtnmax was less in experienced intubators (anaesthetist and residents) than in inexperienced intubators (all other groups) (Student's t test: P = 0.04).
In contrast to the effect on forces exerted in the axial direction, experience proved to have a beneficial effect on the forces in the transverse direction.
本研究调查了插管者经验水平对麦氏喉镜在上颌切牙轴向和横向施加力的影响。
五组不同经验水平的人员(每组15人),包括麻醉科 staff、住院医师、麻醉护士、外科医生和学生,在配备有测量这些力的传感器的插管人体模型上进行一次喉镜检查及随后的插管操作。
46例(61%)朝向颅底的最大横向力(Fmtpmax)在0至10N之间,21例(28%)在10至20N之间,8例(11%)大于或等于20N(最大46.5N)。69例(92%)朝向插管者的横向力最大值(Fmtnmax)在0至10N之间,3例(4%)在10至20N之间,3例(4%)大于或等于20N(最大36.4N)。经验水平与Fmtpmax相关(Spearman检验:P = 0.04,r = 0.24)。此外,经验丰富的插管者(麻醉科 staff和住院医师)的Fmtnmax低于经验不足的插管者(所有其他组)(Student's t检验:P = 0.04)。
与对轴向施加力的影响不同,经验被证明对横向力有有益影响。