Rantanen J, Aro H
Department of Surgery, University of Turku, Finland.
J Orthop Trauma. 1996;10(5):348-50. doi: 10.1097/00005131-199607000-00010.
The Gamma nail, an implant specifically designed for intertrochanteric and subtrochanteric femoral fractures, has been criticized for its high risk of secondary shaft fractures. A modified design, the intramedullary hip screw has recently been introduced to correct this complication. We present a case of mechanical failure of this new implant that occurred in a pathological subtrochanteric fracture. The centering sleeve of the implant became loose and migrated while the head screw penetrated the acetabulum. Refixation was successful.
Gamma钉是一种专门为股骨转子间和转子下骨折设计的植入物,但因其继发股骨干骨折的风险较高而受到批评。最近引入了一种改良设计的髓内髋螺钉来纠正这一并发症。我们报告一例发生在病理性转子下骨折的这种新型植入物机械故障的病例。当头部螺钉穿透髋臼时,植入物的对中套筒松动并移位。再次固定成功。