Suppr超能文献

Development of a clinically applicable high-speed flow cytometer for the isolation of transplantable human hematopoietic stem cells.

作者信息

Sasaki D T, Tichenor E H, Lopez F, Combs J, Uchida N, Smith C R, Stokdijk W, Vardanega M, Buckle A M, Chen B

机构信息

Flow Cytometry Group, SyStemix, Inc., Palo Alto, CA 94304, USA.

出版信息

J Hematother. 1995 Dec;4(6):503-14. doi: 10.1089/scd.1.1995.4.503.

Abstract

A high-speed cell sorter capable of a throughput speed 4-5-fold greater than commercially available systems was developed and evaluated as a processing tool for isolating purified hematopoietic stem cell grafts. The clinical high-speed sorter (CHSS) serves as a single-pass, multiparameter processing tool that provides the means to isolate a highly purified population of cells from starting cell populations with extremely low frequencies. The sorter incorporates environmental barriers to create a sterile environment for cell processing. Monoclonal antibodies and reagents produced under good manufacturing practices (GMP) are used to isolate hematopoietic stem cells by means of the CHSS. Using this technology, the CD34+Thy-1+Lin- hematopoietic stem cell population has been isolated from normal adult bone marrow and mobilized peripheral blood. The sorted cells have been shown to be sterile and viable and to retain hematopoietic function.