使用完全可吸收的自增强聚-L-丙交酯扩张塞和螺钉固定髌腱骨移植的强度。一项在牛尸体上的实验研究。

Strength of the fixation of patellar tendon bone grafts using a totally absorbable self-reinforced poly-L-lactide expansion plug and screw. An experimental study in a bovine cadaver.

作者信息

Tuompo P, Partio E K, Jukkala-Partio K, Pohjonen T, Helevirta P, Rokkanen P

机构信息

Hyvinkää District Hospital, Finland.

出版信息

Arthroscopy. 1996 Aug;12(4):422-7. doi: 10.1016/s0749-8063(96)90035-3.

Abstract

In a bovine cadaver study, bone-tendon-bone graft fixation strength with different graft geometry and fixation devices was measured to evaluate the fixation strength of totally absorbable implants: a 6.0-mm expansion plug and 6.3-mm screw both made of self-reinforced polylactide (SR-PLLA). Comparison was made with 6.5-mm AO cancellous screw. Maximum tensile force to dislodge the bone plug from the bone tunnel was recorded. First, two preliminary tests were performed. In the first test, triangular bone plugs were used (9-mm diameter). The direction of the pull force was parallel to the bone tunnel. The maximum tensile forces were 786 N in femoral insertions and 625 N in tibial insertions, mean. After this, we evaluated the influence of change in the pullout direction. In the second test, a circular bone plug was used with no fixation but the direction of the pull force was parallel to the tibial or femoral axis and the bone plug (10-mm diameter) was in a 30 degrees to 40 degrees angle to the direction of the pull force and it was compressed to the tunnel (9-mm diameter). The maximum tensile forces were 783 N in femoral insertions and 695 N in tibial insertions, mean. In the final third test, we used a curved saw in harvesting the graft. This made a half-circular bone block with a diameter of 12 mm. The maximum tensile force to dislodge the bone plug from the bone tunnel was recorded and the pull force was in a 30 degrees to 40 degrees angle to the tunnel. The results were evaluated with Student's t-test and Mann-Whitney U-test. With the AO screw, the maximum tensile force to dislodge the bone plug from the bone tunnel was 2,113 +/- 407 N (mean +/- standard deviation) and it was better than the fixation strength of the SR-PLLA expansion plug, 1,379 +/- 328 N (P = .009, t-test) and better than the fixation strength of SR-PLLA screw, 1,454 +/- 230 N (P = .007, t-test). However, the maximum tensile force of both SR-PLLA implants in all measurements in the third test were above 1,100 N and it seems that the initial strength of totally absorbable implants is enough for the clinical use.

摘要

在一项牛尸体研究中,测量了不同移植物几何形状和固定装置的骨-肌腱-骨移植物固定强度,以评估完全可吸收植入物的固定强度:一个6.0毫米的扩张塞和一个6.3毫米的螺钉,均由自增强聚丙交酯(SR-PLLA)制成。与6.5毫米的AO松质骨螺钉进行了比较。记录了将骨塞从骨隧道中拔出的最大拉力。首先,进行了两项初步测试。在第一次测试中,使用了三角形骨塞(直径9毫米)。拉力方向与骨隧道平行。股骨植入的平均最大拉力为786牛,胫骨植入的平均最大拉力为625牛。在此之后,我们评估了拔出方向变化的影响。在第二次测试中,使用了一个无固定的圆形骨塞,但拉力方向与胫骨或股骨轴线平行,骨塞(直径10毫米)与拉力方向成30度至40度角,并被压入隧道(直径9毫米)。股骨植入的平均最大拉力为783牛,胫骨植入的平均最大拉力为695牛。在最后的第三次测试中,我们在采集移植物时使用了弯锯。这制作了一个直径12毫米的半圆形骨块。记录了将骨塞从骨隧道中拔出的最大拉力,拉力与隧道成30度至40度角。结果用学生t检验和曼-惠特尼U检验进行评估。使用AO螺钉时,将骨塞从骨隧道中拔出的最大拉力为2113±407牛(平均值±标准差),它优于SR-PLLA扩张塞的固定强度,即1379±328牛(P = 0.009,t检验),也优于SR-PLLA螺钉的固定强度,即1454±230牛(P = 0.007,t检验)。然而,在第三次测试的所有测量中,两种SR-PLLA植入物的最大拉力均高于1100牛,似乎完全可吸收植入物的初始强度足以用于临床。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验