Sumiya T, Suzuki Y, Kasahara T
Rosai Rehabilitation Engineering Center, Nagoya, Japan.
Prosthet Orthot Int. 1996 Aug;20(2):129-31. doi: 10.3109/03093649609164430.
A device was developed to measure the dorsi- and plantar flexion moment of plastic ankle-foot orthoses when deflected. It is operated by manually controlling a lever which is used to apply a nearly static force. Various orthoses can be classified according to the characteristics of the correcting force measured by this device. Simplicity and high reproducibility are the major advantages. However, to obtain measurements approximating the characteristics of orthoses under wearing conditions its use is restricted to orthoses made of low-viscosity materials.
开发了一种装置,用于测量塑料踝足矫形器在发生偏转时的背屈和跖屈力矩。该装置通过手动控制一个杠杆来操作,该杠杆用于施加近乎静态的力。各种矫形器可以根据该装置测量的矫正力特性进行分类。简单性和高再现性是主要优点。然而,为了获得接近佩戴条件下矫形器特性的测量结果,其使用仅限于由低粘度材料制成的矫形器。