Mishima M, Kawakami K, Fukunaga T, Sugiura N, Hirai T, Oku Y, Fukui M, Chin K, Ohi M, Kuno K
Department of Clinical Physiology, Kyoto University, Japan.
Front Med Biol Eng. 1996;7(3):163-75.
The accuracy and validity of a body box system which was developed for the measurement of airway and chest wall impedance in small animals such as canines was evaluated. Input impedance was calibrated using a resistance tube and the mouth flow sensing system was designed to be as symmetrical as possible such that the minimal common mode rejection ratio (CMRR) from 10 to 40 Hz was greater than 60 dB. The validity of this system using the resistance tube and inertant models indicated that the average error ratio for the input impedance and for the flow transfer function between the mouth and chest flow was within 3 and 0.5%, respectively. The location of the box pressure port near to the oscillatory flow inlet was shown to potentially give rise to errors in the measurement of the flow transfer factor due to the direct effects of the inflow. A distance of 60 cm was adopted in the present system, which proved to be sufficient to avoid this effect. It was concluded that the body box system which is described for small animals in the present study is appropriate for obtaining separate and accurate airway and tissue impedance data.
对一种为测量犬类等小动物气道和胸壁阻抗而开发的体箱系统的准确性和有效性进行了评估。使用电阻管对输入阻抗进行校准,并且口部流量传感系统设计得尽可能对称,使得10至40Hz的最小共模抑制比(CMRR)大于60dB。使用电阻管和惰性模型对该系统的有效性表明,输入阻抗和口部与胸部流量之间的流量传递函数的平均误差率分别在3%和0.5%以内。结果表明,由于流入的直接影响,箱压力端口靠近振荡流入口的位置可能会在流量传递因子的测量中产生误差。本系统采用60cm的距离,事实证明这足以避免这种影响。得出的结论是,本研究中描述的用于小动物的体箱系统适合获取单独且准确的气道和组织阻抗数据。