Poor in vivo reproducibility of gastric intramucosal pH determined by saline-filled balloon tonometry.

作者信息

Oud L, Kruse J A

机构信息

Division of Pulmonary/Critical Care Medicine, Wayne State University School of Medicine, Detroit, Michigan, USA.

出版信息

J Crit Care. 1996 Sep;11(3):144-50. doi: 10.1016/s0883-9441(96)90011-8.

Abstract

PURPOSE

The reproducibility of gastric intramucosal pH (pHi) determinations by saline-filled balloon tonometry has not been assessed adequately. We examined the agreement of pHi obtained using pairs of tonometry catheters under various conditions in a canine model of global hypoxia.

METHODS

Two gastric balloon tonometry catheters were inserted into each of 16 anesthetized dogs. Cimetidine was administered to six of the animals (group 1). Hypoxia was induced for 30 minutes by reducing the F1O2 to 0.08. pHi was determined under basal conditions, at completion of hypoxia, and during the posthypoxic period.

RESULTS

The mean bias of the paired pHi determinations was 0.03 for group 1 and 0.00 for group 2 (P = NS). The corresponding 95% limits of agreement spanned 0.24 and 0.28 pH units, respectively. Agreement of pHi determinations between tonometry catheters was not significantly affected by varying of the equilibration period, by administration of cimetidine, or by the presence of posthypoxic conditions.

CONCLUSIONS

Clinically important disagreement was observed in simultaneous pHi determinations obtained using separate gastric balloon tonometry catheters exposed to identical in vivo conditions. Poor reproducibility of individual pHi determinations may limit the accuracy of pHi obtained by gastric balloon tonometry in the clinical setting.

摘要

目的

用充生理盐水的气囊张力测定法测定胃黏膜内pH值(pHi)的可重复性尚未得到充分评估。我们在犬类全脑缺氧模型中,研究了在不同条件下使用一对张力测定导管所测得的pHi的一致性。

方法

将两根胃气囊张力测定导管插入16只麻醉犬的每只体内。给其中6只动物(第1组)使用西咪替丁。通过将吸入氧分数降至0.08诱导缺氧30分钟。在基础条件下、缺氧结束时以及缺氧后阶段测定pHi。

结果

第1组配对pHi测定的平均偏差为0.03,第2组为0.00(P = 无显著性差异)。相应的95%一致性界限分别为0.24和0.28个pH单位。张力测定导管之间pHi测定的一致性不受平衡期变化、西咪替丁给药或缺氧后条件的影响。

结论

在相同体内条件下,使用单独的胃气囊张力测定导管同时测定pHi时,观察到具有临床意义的差异。个体pHi测定的可重复性差可能会限制胃气囊张力测定法在临床环境中所测pHi的准确性。