Müller B, Hasman A, Blom J A
Department of Medical Electrical Engineering, Eindhoven University of Technology, The Netherlands.
Int J Biomed Comput. 1996 Aug;42(3):165-79. doi: 10.1016/0020-7101(96)01210-x.
In an earlier study an approach was described to generate intelligent alarm systems for monitoring ventilation of patients via mathematical simulation and machine learning. However, ventilator settings were not varied. In this study we investigated whether an alarm system could be created with which a satisfactory classification performance could be obtained under a wide variety of ventilator settings, by varying inspiratory to expiratory time (I:E) ratio, tidal volume and respiratory rate. In a first experiment three patient data sets were modeled, each with a different I:E ratio. A part of each data set was used to construct an alarm system for each I:E ratio. The remaining part was used to test the performance of the alarm systems. The three training sets were also combined to construct one alarm system, which was tested with the three test sets. Finally, all alarm systems were tested with data generated by a patient simulator. Similar experiments were performed for the tidal volume and the respiratory rate. It was concluded that an optimally functioning alarm system should contain a library of rule sets, one for each set of ventilator settings. A second best alternative is to take all possible settings into consideration when constructing the training set. Classification performance of the trees that were trained with multiple ventilator settings ranged from 98 to 100% for all test sets. When tested with the independent patient simulator data the classification performance of these trees ranged from 80 to 100%.
在一项早期研究中,描述了一种通过数学模拟和机器学习生成用于监测患者通气的智能报警系统的方法。然而,通气机设置并未改变。在本研究中,我们调查了是否可以创建一种报警系统,通过改变吸气与呼气时间(I:E)比率、潮气量和呼吸频率,在各种通气机设置下获得令人满意的分类性能。在第一个实验中,对三个患者数据集进行建模,每个数据集具有不同的I:E比率。每个数据集的一部分用于为每个I:E比率构建一个报警系统。其余部分用于测试报警系统的性能。这三个训练集也合并起来构建一个报警系统,并用这三个测试集进行测试。最后,所有报警系统都用患者模拟器生成的数据进行测试。对潮气量和呼吸频率进行了类似的实验。得出的结论是,一个功能最佳的报警系统应该包含一个规则集库,每个通气机设置集对应一个规则集。次优的选择是在构建训练集时考虑所有可能的设置。对于所有测试集,用多种通气机设置训练的树的分类性能在98%至100%之间。当用独立的患者模拟器数据进行测试时,这些树的分类性能在80%至100%之间。