Evans J M, Hogg M I, Rosen M
Br J Anaesth. 1977 Aug;49(8):761-4. doi: 10.1093/bja/49.8.761.
The performance of an infra-red analyser (Beckman LB-2) in sampling alveolar gas at ventilatory frequencies and volumes occuring in the human neonate has been examined. The optimum sampling flow rate was found to be 300 ml/min; at this flow the 90% response time of the analyser-catheter system was 140 ms. Correlation of estimated alveolar PCO2 and arterial PCO2 was performed in the rabbit which has a ventilatory rate and tidal volume similar to those of the human neonate. There was a good correlation between maximum end-expired PCO2 (PE'CO2) and arterial PCO2 over a wide range of ventilatory rates (10-100 b.p.m.); maximum PE'CO2 was a better index than mean PE'CO2 or alveolar plateau PE'CO2.
已对一台红外分析仪(贝克曼LB - 2型)在采集人类新生儿出现的通气频率和潮气量下的肺泡气时的性能进行了检测。发现最佳采样流速为300毫升/分钟;在此流速下,分析仪 - 导管系统的90%响应时间为140毫秒。在通气频率和潮气量与人类新生儿相似的家兔身上进行了估计肺泡PCO2与动脉PCO2的相关性研究。在很宽的通气频率范围(10 - 100次/分钟)内,最大呼气末PCO2(PE'CO2)与动脉PCO2之间存在良好的相关性;最大PE'CO2比平均PE'CO2或肺泡平台PE'CO2是更好的指标。