Diederichs C G, Engelke W G, Richter B, Hermann K P, Oestmann J W
Department of Radiology, Georg-August-University of Goettingen, Germany.
AJNR Am J Neuroradiol. 1996 Oct;17(9):1758-60.
We evaluated the feasibility of performing preoperative spiral CT of the maxilla and mandible with a radiation dose similar to that used for conventional panoramic radiography. The skin entrance doses of radiation used for spiral CT (collimation, 1 mm; pitch, 2; tube voltage, 80 kV; tube current, 40 mA) and for panoramic radiography (75 kV, 8 mA, 15 seconds) were measured in one patient by using thermoluminescent dosimeter chips. Results were 0.56 +/- 0.06 mGy for CT and 0.59 +/- 0.04 mGy for radiography. Image quality was adequate for preoperative implant planning. Spiral CT of the mandible and maxilla may therefore be feasible with a radiation dose of similar magnitude as that used for conventional panoramic radiography.
我们评估了进行上颌骨和下颌骨术前螺旋CT检查的可行性,其辐射剂量与传统全景X线摄影所用剂量相似。通过使用热释光剂量计芯片,在一名患者身上测量了螺旋CT(准直1mm;螺距2;管电压80kV;管电流40mA)和全景X线摄影(75kV,8mA,15秒)的皮肤入射辐射剂量。CT检查结果为0.56±0.06mGy,X线摄影结果为0.59±0.04mGy。图像质量对于术前种植计划而言足够。因此,下颌骨和上颌骨的螺旋CT检查在与传统全景X线摄影相似的辐射剂量下可能是可行的。