两种双能X线吸收仪对年轻男性身体成分和骨矿物质测量结果的比较。

Comparison of body composition and bone mineral measurements from two DXA instruments in young men.

作者信息

Modlesky C M, Lewis R D, Yetman K A, Rose B, Rosskopf L B, Snow T K, Sparling P B

机构信息

Department of Foods and Nutrition, University of Georgia, Athens, USA.

出版信息

Am J Clin Nutr. 1996 Nov;64(5):669-76. doi: 10.1093/ajcn/64.5.669.

Abstract

The purpose of this study was to compare body-composition and whole-body, lumbar spine (LS), femoral neck (FN), trochanter, and Ward's triangle (WT) bone mineral measurements by using the Hologic QDR 1000W (DXAH) and the Lunar DPX-L (DXAL) dual-energy X-ray absorptiometry instruments. In addition, the ability of conversion equations to predict DXAH data from DXAL data were tested. Thirteen healthy young adult males (aged 22.2 +/- 3.6 y, 177.4 +/- 5.3 cm in height, and 72.7 +/- 9.6 kg in weight) were scanned on the same day by using DXAH and DXAL. Whereas measured body mass was not different (P > 0.05) between machines, whole-body fat mass [DXAH-DXAL (DXAdiff) = 1152 +/- 1395 g], percentage fat (DXAdiff = 1.5 +/- 1.7% of body mass), bone mineral density (BMD; DXAdiff = 0.016 +/- 0.023 g/cm2), and bone mineral content (BMC; DXAdiff = 316 +/- 50 g) were lower and whole-body fat-free soft tissue (FFST; DXAdiff = 1781 +/- 1859 g) was higher with DXAH than with DXAL. Lower fat mass (DXAdiff = 2145 +/- 855 g) and BMC (DXAdiff = 216 +/- 36 g) and higher FFST (DXAdiff = 1966 +/- 943 g) in the trunk were primarily responsible for the whole-body differences. Lower BMD and BMC values were found for LS (DXAdiff = 0.145 +/- 0.038 g/cm2 and 3 +/- 2 g, respectively), trochanter (DXAdiff = 0.100 +/- 0.044 g/cm2 and 1.7 +/- 1.0 g), and WT (DXAdiff = 0.195 +/- 0.061 g/cm2 and 1.93 +/- 0.51 g) with DXAH compared with DXAL. DXAH BMD of FN was also lower (DXAdiff = 0.141 +/- 0.032 g/cm2) than with DXAL. Only DXAH whole-body BMC and LS BMD were accurately predicted from DXAL with conversion equations. Predicted DXAH FN BMD was significantly lower than the actual DXAH value (P < or = 0.05), whereas the discrepancy between DXAH and DXAL actually increased for whole-body percentage fat and BMD (DXAdiff = 6.6 +/- 1.3% body mass and 0.020 +/- 0.025 g/cm2). In conclusion, lower whole-body fat mass, percentage fat, and BMC, and higher whole-body FFST with DXAH were due primarily to measurement differences in the trunk. Whereas conversion equations accurately predicted DXAH whole-body BMC and LS BMD from DXAL measurements in young adult males, they did not accurately predict DXAH, percentage fat, and BMD of the whole body and FN BMD.

摘要

本研究的目的是使用Hologic QDR 1000W(DXAH)和Lunar DPX-L(DXAL)双能X线吸收仪比较身体成分以及全身、腰椎(LS)、股骨颈(FN)、大转子和沃德三角(WT)的骨矿物质测量值。此外,还测试了转换方程从DXAL数据预测DXAH数据的能力。13名健康年轻成年男性(年龄22.2±3.6岁,身高177.4±5.3厘米,体重72.7±9.6千克)在同一天使用DXAH和DXAL进行扫描。虽然两台仪器测量的体重没有差异(P>0.05),但与DXAL相比,DXAH测量的全身脂肪量[DXAH - DXAL(DXAdiff)=1152±1395克]、脂肪百分比(DXAdiff =体重的1.5±1.7%)、骨矿物质密度(BMD;DXAdiff =0.016±0.023克/平方厘米)和骨矿物质含量(BMC;DXAdiff =316±50克)较低,而全身去脂软组织(FFST;DXAdiff =1781±1859克)较高。躯干中较低的脂肪量(DXAdiff =2145±855克)和BMC(DXAdiff =216±36克)以及较高的FFST(DXAdiff =1966±943克)是造成全身差异的主要原因。与DXAL相比,DXAH测量的LS(DXAdiff分别为0.145±0.038克/平方厘米和3±2克)、大转子(DXAdiff =0.100±0.044克/平方厘米和1.7±1.0克)和WT(DXAdiff =0.195±0.061克/平方厘米和1.93±0.51克)的BMD和BMC值较低。DXAH测量的FN的BMD也低于DXAL(DXAdiff =0.141±0.032克/平方厘米)。只有DXAH的全身BMC和LS的BMD能够通过转换方程从DXAL数据中准确预测。预测的DXAH的FN的BMD显著低于实际DXAH值(P≤0.05),而全身脂肪百分比和BMD的DXAH与DXAL之间的差异实际上增加了(DXAdiff =体重的6.6±1.3%和0.020±0.025克/平方厘米)。总之,DXAH测量的较低的全身脂肪量、脂肪百分比和BMC以及较高的全身FFST主要是由于躯干测量差异所致。虽然转换方程能够从年轻成年男性的DXAL测量值中准确预测DXAH的全身BMC和LS的BMD,但它们不能准确预测DXAH的全身脂肪百分比、BMD以及FN的BMD。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验