The Björk-Shiley Delrin tilting disc heart valve: historical perspective, design and need for scientific analyses after 25 years.

作者信息

Wieting D W

机构信息

Shiley Incorporated, Irvine, Calífornia 92606, USA.

出版信息

J Heart Valve Dis. 1996 Aug;5 Suppl 2:S157-68.

PMID:8905516
Abstract

The Björk-Shiley Delrin (BSD) tilting disc heart valve was first used clinically in 1969. It is estimated that up to 24,000 BSD heart valves were implanted between 1969 and 1981, of which 7,000 may still be implanted in patients. The BSD valve provided a low-profile, quiet prosthesis with excellent hemodynamics. There have been only two reports of mechanical failure due to inlet strut fracture. Recent reports of wear of the Delrin disc combined with wear as observed on explanted BSD valves after 17-20 years in patients stimulated the formation of a Scientific Advisory Panel by Shiley to review the status of the BSD heart valve and make recommendations regarding patients living with BSD heart valves. The studies presented in this Supplement of The Journal of Heart Valve Disease summarize some of the scientific investigations conducted by Shiley and those recommended by this Panel. The material reviewed by the Panel included: (i) clinical publications on BSD heart valves; (ii) seventy-three explanted BSD heart valves returned to Shiley over the past 25 years; (iii) BSD valves subjected to accelerated wear tests; (iv) unimplanted BSD heart valves in Shiley's inventory; (v) in vitro studies of static leakage and pulsatile regurgitation of explanted and unimplanted "control' BSD heart valves; (vi) two Christiansen hip prostheses explanted from patients after five and 13 years; (vii) implant data for the BSD heart valve from the Karolinska Hospital, Stockholm, Sweden; and (viii) test reports for studies and analyses conducted by Shiley for the BSD heart valve. This article gives an overview of and introduction to the studies of the BSD heart valve presented in the following papers.

摘要

相似文献

1
The Björk-Shiley Delrin tilting disc heart valve: historical perspective, design and need for scientific analyses after 25 years.
J Heart Valve Dis. 1996 Aug;5 Suppl 2:S157-68.
2
Summary and recommendations.总结与建议。
J Heart Valve Dis. 1996 Aug;5 Suppl 2:S246-8.
3
Clinical experience with the Björk-Shiley Delrin tilting disc heart valve.Björk-Shiley Delrin倾斜碟片心脏瓣膜的临床经验。
J Heart Valve Dis. 1996 Aug;5 Suppl 2:S169-77.
4
Surface and edge wear of Björk-Shiley Delrin heart valve discs.
J Heart Valve Dis. 1996 Aug;5 Suppl 2:S190-205.
5
A comparison of wear patterns in vivo and in vitro for Björk-Shiley Delrin heart valve discs.
J Heart Valve Dis. 1996 Aug;5 Suppl 2:S206-15.
6
Evaluation of disc retention of Björk-Shiley Delrin heart valves.
J Heart Valve Dis. 1996 Aug;5 Suppl 2:S243-5.
7
Variability of regurgitation in Björk-Shiley mitral valves and relationship to disc occluder design: an in vitro two-dimensional color-Doppler flow mapping study.比约克-希利二尖瓣反流的变异性及其与碟片封堵器设计的关系:一项体外二维彩色多普勒血流图研究
J Heart Valve Dis. 1996 Aug;5 Suppl 2:S178-83.
8
Theory of wear as related to the Björk-Shiley Delrin disc heart valve.
J Heart Valve Dis. 1996 Aug;5 Suppl 2:S216-28.
9
Ex-vivo characterization of three Björk-Shiley Delrin heart valves.
J Heart Valve Dis. 2008 May;17(3):325-31.
10
Structural analysis of the Björk-Shiley Delrin heart valve occluder.比约克-希利Delrin心脏瓣膜封堵器的结构分析。
J Heart Valve Dis. 1996 Aug;5 Suppl 2:S229-37.

引用本文的文献

1
Design Variation, Implantation, and Outcome of Transcatheter Mitral Valve Prosthesis: A Comprehensive Review.经导管二尖瓣假体的设计变异、植入及结果:一项综述
Front Cardiovasc Med. 2022 Feb 24;8:782278. doi: 10.3389/fcvm.2021.782278. eCollection 2021.
2
Design, Manufacturing Technology and In-Vitro Evaluation of Original, Polyurethane, Petal Valves for Application in Pulsating Ventricular Assist Devices.用于脉动式心室辅助装置的原创聚氨酯花瓣阀的设计、制造技术及体外评估
Polymers (Basel). 2020 Dec 15;12(12):2986. doi: 10.3390/polym12122986.
3
Replacement of a Björk-Shiley Delrin aortic valve still functioning after 25 years.
更换使用25年后仍在正常工作的比约克-希利德尔林主动脉瓣。
Tex Heart Inst J. 2004;31(3):303-5; discussion 305.
4
Concentric wear of the Delrin disc in a Bjork-Shiley heart valve prosthesis: report of two cases.
Surg Today. 1999;29(10):1115-9. doi: 10.1007/s005950050656.