Naus A J, Kuppens P S, Borst A
Clin Chem. 1977 Sep;23(9):1624-7.
A method for calculating radioimmunoassay standard curves, based on the theory of Ekins et al., is described. Because a four-parameter model is used, nonlinear standard curves are the result. The calibration curve is fitted to the measured standard points by means of a weighted least-squares method. The program based on this model can be easily processed on a desk-top calculator. For all 250 runs of six different assays, very good standard curves could be obtained. The mean deviation between the concentrations of the standard points and the corresponding calculated values was about 6%. In 26% of the cases it could be shown that the model we describe gave significantly better results than did two simpler ones.
本文描述了一种基于埃金斯等人理论的放射免疫分析标准曲线计算方法。由于使用了四参数模型,得到的是非线性标准曲线。校准曲线通过加权最小二乘法拟合到测量的标准点上。基于此模型的程序可以很容易地在台式计算器上进行处理。对于六种不同分析方法的所有250次运行,都能获得非常好的标准曲线。标准点浓度与相应计算值之间的平均偏差约为6%。在26%的情况下,可以证明我们所描述的模型比另外两种更简单的模型给出的结果明显更好。