Nobbs B T, Smith J M, Walker A W
Clin Chim Acta. 1977 Sep 1;79(2):391-7. doi: 10.1016/0009-8981(77)90434-x.
Enzymic procedures for the determination of plasma cholesterol, using cholesterol esterase and cholesterol oxidase, have been adapted to the Vickers D-300, Vickers M,-300, and Vitatron AKES discrete analysers. The results obtained by these methods have been compared to those obtained by manual and continuous flow Liebermann-Burchard methods. The enzymic methods were found to be accurate, precise and of adequate sensitivity.
使用胆固醇酯酶和胆固醇氧化酶测定血浆胆固醇的酶法已被应用于Vickers D - 300、Vickers M - 300和Vitatron AKES离散分析仪。这些方法所获得的结果已与通过手工和连续流动的Liebermann - Burchard方法所获得的结果进行了比较。发现酶法准确、精密且灵敏度足够。