Abstract

We describe a rapid, kinetic, fixed-time method for determining serum total cholesterol by use of cholesterol esterase, cholesterol oxidase, and the indicator reaction with peroxidase, 4-aminophenazone, and phenol. On addition of the competitive inhibitor 3,4-dichlorophenol the Michaelis constant of cholesterol oxidase is apparently increased, which extends the linear relation between absorbance change and cholesterol concentration to 20.7 to 25.9 mmol/L, depending on the analyzer being used. For calibration, a single standard is used. Total analysis time is in the range of 80 to 210 s. Incubation temperature is 25 degrees C or 37 degrees C. The single-reagent procedure has been adapted to three different centrifugal analyzers and to the Eppendorf ACP 5040 analyzer. It yields precise and accurate results and is insensitive to potential interferences.

摘要

我们描述了一种快速、动力学、定时的方法,用于通过使用胆固醇酯酶、胆固醇氧化酶以及与过氧化物酶、4-氨基苯乙酮和苯酚的指示反应来测定血清总胆固醇。加入竞争性抑制剂3,4-二氯苯酚后,胆固醇氧化酶的米氏常数明显增加,这使得吸光度变化与胆固醇浓度之间的线性关系扩展至20.7至25.9 mmol/L,具体取决于所使用的分析仪。校准采用单一标准品。总分析时间在80至210秒范围内。孵育温度为25℃或37℃。该单试剂程序已适用于三种不同的离心分析仪以及艾本德ACP 5040分析仪。它能产生精确且准确的结果,并且对潜在干扰不敏感。