Tracht S E, Cruz L, Stobba-Wiley C M, Sweedler J V
Department of Chemistry, University of Illinois, Urbana 61801, USA.
Anal Chem. 1996 Nov 15;68(22):3922-7. doi: 10.1021/ac9603901.
A capillary electrophoresis (CE) postcolumn radionuclide detector has been developed that uses a commercial phosphor-imaging detector and has been optimized for low-energy beta emitters. Eluant from the separation capillary is deposited on a membrane. Emission from radioactive analytes on the membrane is integrated using the phosphor-imaging system for 10-72 h. Results from the phosphor-imaging system are converted to conventional electropherograms. Modifications to a prior postcolumn CE deposition system have been accomplished by adding a buffer makeup capillary; this increases the electrolyte pH range and improves reproducibility. The limit of detection (LOD) for 35S-labeled analytes is 0.13 amol (8.7 pM or 0.007 Bq), while the LOD for 32P-labeled analytes is 4.9 zmol (0.33 pM or 0.002 Bq), with a linear range for 35S-Met from 1.5 amol to 1.5 fmol.
已开发出一种毛细管电泳(CE)柱后放射性核素检测器,它使用商用磷光成像检测器,并针对低能β发射体进行了优化。分离毛细管中的洗脱液沉积在膜上。使用磷光成像系统对膜上放射性分析物的发射进行10至72小时的积分。磷光成像系统的结果被转换为传统的电泳图谱。通过添加缓冲液补充毛细管对先前的柱后CE沉积系统进行了改进;这扩大了电解质pH范围并提高了重现性。35S标记分析物的检测限(LOD)为0.13 amol(8.7 pM或0.007 Bq),而32P标记分析物的LOD为4.9 zmol(0.33 pM或0.002 Bq),35S-蛋氨酸的线性范围为1.5 amol至1.5 fmol。