[Interactive stereotaxic interventions in superconducting, open 0.5-Tesla MRI tomography].

作者信息

Steiner P, Schoenenberger A W, Penner E A, Erhart P, Debatin J F, von Schulthess G K, Kacl G M

机构信息

Departement Medizinische Radiologie, Universitätsspital Zürich.

出版信息

Rofo. 1996 Sep;165(3):276-80. doi: 10.1055/s-2007-1015755.

Abstract

PURPOSE

Evaluation of an interactive, stereotactic biopsy device integrated in an open superconductive 0.5 tesla MR-scanner with a vertical gap.

MATERIAL AND METHODS

In addition to "in-vitro" experiments performed on a plexiglas phantom with holes of varying diameters (5-20 mm) biopsies on eleven patients (7 women, 4 men; average age 55 years) were performed in the interventional MR-system. Lesions in the abdomen (n = 6), muscle (n = 1), thyroid (n = 3) and breast (n = 1) were targeted with 18-20 G aspiration biopsy needles of 5-15 cm length. The intervention was interactively guided by a fast T1-weighted 2-D gradient echo sequence.

RESULTS

All of the 15 and 20 mm holes of the phantom, but only 83% of the 10 and 5 mm holes were hit. No complications occurred during the MR-guided patient procedures. All lesions (mean size 3.5 cm, distance from the skin 2 cm to 10 cm) were biopsied successfully. The fast image acquisition in combination with the stereotactic technique enables interactive control of the needle.

CONCLUSIONS

Stereotactic, interactively controlled biopsies in the interventional MR are technical feasible. However, the range of meaningful indications for MR-guided biopsies is limited.