Elsner M, Peifer A, Kasper W
Medizinische Klinik I (Dep. of Cardiology), St. Josefs Hospital, Wiesbaden, Germany.
Cathet Cardiovasc Diagn. 1996 Nov;39(3):271-6. doi: 10.1002/(SICI)1097-0304(199611)39:3<271::AID-CCD13>3.0.CO;2-I.
Intracoronary loss of balloon-mounted stents prior to deployment is a rare event but may result in clinically relevant cardiac ischemia or peripheral embolization during rescue attempts. In a consecutive series of nearly 400 patients undergoing elective stenting, we encountered six instances of intracoronary misplacement. We describe four cases of successful percutaneous retrieval using a 2 mm closed loop nitinol device ("Microsnare"). Our proposed method of extraction seems to hold advantages over those previously described.