通过内标选择提高甲基苯丙胺气相色谱-质谱联用分析中低浓度下的离子质量比性能。

Improving ion mass ratio performance at low concentrations in methamphetamine GC-MS assay through internal standard selection.

作者信息

Urry F M, Kushnir M, Nelson G, McDowell M, Jennison T

机构信息

Associated Regional and University Pathologists, Inc., Salt Lake City, Utah 84108, USA.

出版信息

J Anal Toxicol. 1996 Nov-Dec;20(7):592-5. doi: 10.1093/jat/20.7.592.

Abstract

In federally regulated drug testing, laboratories must identify and quantitate drugs and their breakdown products by gas chromatography-mass spectrometry (GC-MS) to a concentration that is at least 60% below the cutoff concentration for reconfirmation purposes. Use of methamphetamine-d5 as an internal standard in routine testing with derivatization by HFBA was found to contribute m/z 91 and 118 ions to the same ions from the nondeuterated methamphetamine in the specimen. This resulted in poor chromatography and occasionally caused the 91/254 and 118/254 ion mass ratios to exceed the +20% acceptance limit established in the calibration process at low concentrations. The analogues methamphetamine-d8 and -d11 were evaluated for contributions to the nondeuterated methamphetamine ion fragments. Methamphetamine-d8 produced m/z 91 and 118 ions, but in less abundance than methamphetamine-d5. Methamphetamine-d11 was found to produce little or no detectable m/z 91 or 118. Replacing methamphetamine-d5 with methamphetamine-d11 eliminates this problem and allows the assay to consistently produce ion mass ratios and acceptable chromatography sufficient for identifying and quantitating methamphetamine at 60% below the 500-ng/mL cutoff concentration.

摘要

在联邦监管的药物检测中,实验室必须通过气相色谱 - 质谱联用仪(GC - MS)识别并定量药物及其分解产物,使其浓度至少比用于重新确认目的的截断浓度低60%。发现在使用HFBA衍生化的常规检测中,将甲基苯丙胺 - d5用作内标会使标本中未氘代甲基苯丙胺产生的相同离子增加m/z 91和118离子。这导致色谱效果不佳,并且在低浓度时偶尔会使91/254和118/254离子质量比超过校准过程中设定的 +20% 接受限。对甲基苯丙胺 - d8和 - d11类似物对未氘代甲基苯丙胺离子碎片的贡献进行了评估。甲基苯丙胺 - d8产生m/z 91和118离子,但丰度低于甲基苯丙胺 - d5。发现甲基苯丙胺 - d11几乎不产生或不产生可检测到的m/z 91或118。用甲基苯丙胺 - d11替代甲基苯丙胺 - d5可消除此问题,并使该测定法能够始终产生足够用于识别和定量甲基苯丙胺的离子质量比和可接受的色谱图,浓度比500 - ng/mL截断浓度低60%。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验