The ocular blood flow tonograph: a new instrument for the measurement of intraocular pressure in rabbits.

作者信息

Chidlow G, Nash M S, Crowhurst C, Bron A J, Osborne N N

机构信息

Nuffield Laboratory of Ophthalmology, University of Oxford, U.K.

出版信息

Exp Eye Res. 1996 Oct;63(4):463-9. doi: 10.1006/exer.1996.0136.

Abstract

The Ocular Blood Flow Tonograph (OBFT) and Digilab Pneumatonograph (PTG) were calibrated for use on New Zealand White rabbits using a pressure manometer and performing tonometry at varying known pressures. The calibration experiments showed that a good correlation exists between OBFT measurements and manometric readings of IOP. Moreover, no difference between open and closed stopcock measurements were observed. Corrective equations for conversion of OBFT readings to mm Hg are documented. The OBFT was compared with the PTG and the results showed there was no significant difference between the IOP measured by either instrument in a group of rabbits displaying a range of different IOPs. Regression analysis of the data gave a slope of 0.88 and a correlation co-efficient of 0.78. The OBFT proved easier to use than the PTG under restricted lighting conditions. The OBFT was also used to measure the circadian variation in IOP and to measure the IOP after the topical application of prostaglandin E2 and various serotonergic agonists. The combined data demonstrated that the OBFT is an accurate and reliable tonometer for the measurement of IOP in rabbits.

摘要

使用压力计对眼血流眼压描记仪(OBFT)和Digilab气动眼压描记仪(PTG)进行校准,以便在新西兰白兔身上使用,并在不同的已知压力下进行眼压测量。校准实验表明,OBFT测量值与眼压计测量的眼压读数之间存在良好的相关性。此外,未观察到打开和关闭旋塞测量之间的差异。记录了将OBFT读数转换为毫米汞柱的校正方程。将OBFT与PTG进行比较,结果表明,在一组显示不同眼压范围的兔子中,两种仪器测量的眼压之间没有显著差异。对数据的回归分析得出斜率为0.88,相关系数为0.78。在光线受限的条件下,OBFT比PTG更易于使用。OBFT还用于测量眼压的昼夜变化,以及局部应用前列腺素E2和各种血清素能激动剂后的眼压。综合数据表明,OBFT是一种准确可靠的眼压计,可用于测量兔子的眼压。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验