Suppr超能文献

在猪模型中对胸腔内人工肺进行测试。

Testing of an intrathoracic artificial lung in a pig model.

作者信息

Cook K E, Makarewicz A J, Backer C L, Mockros L F, Przybylo H J, Crawford S E, Hernandez J M, Leonard R J, Mavroudis C

机构信息

Department of Biomedical Engineering, Northwestern University, Evanston, Illinois, USA.

出版信息

ASAIO J. 1996 Sep-Oct;42(5):M604-9. doi: 10.1097/00002480-199609000-00059.

Abstract

A low input impedance, intrathoracic artificial lung is being developed for use in acute respiratory failure or as a bridge to transplantation. The device uses microporous, hollow fibers in a 0.74 void fraction, 1.83 m2 surface area bundle. The bundle is placed within a thermoformed polyethylene terephthalate glucose modified housing with a gross volume of 800 cm3. The blood inlet and outlet are 18 mm inner diameter vascular grafts. Between the inlet graft and the device is a 1 inch inner diameter, thin-walled, latex tubing compliance chamber. These devices were implanted in Yorkshire pigs via median sternotomy with an end to side anastomosis to the pulmonary artery and left atrium. The distal pulmonary artery was occluded to divert the right ventricular output to the device. Pigs 1 and 2 were supported fully for 24 hrs and then killed. Pig 3 was supported partially for 20 hrs and died from bleeding complications. The first implant, in a 55 kg male pig, transferred an average of 176 ml/min +/- 42.4 of O2 and 190 ml/min +/- 39.7 of CO2 with an average blood flow rate of 2.71/min +/- 0.46. The normalized average right ventricular output power, Pn, was 0.062 W/(L/min) +/- 0.0082, and the average device resistance, R, was 3.5 mmHg/(L/min) +/- 0.62. The second implant, in a 60 kg male pig, transferred an average of 204 ml/min +/- 22.5 of O2 and 242 ml/min +/- 17.2 of CO2 with an average blood flow rate of 3.7 L/min +/- 0.45, Pn of 0.064 W/(L/min) +/- 0.0067, and R of 4.3 mmHg/(L/min) +/- 0.89. The third implant, in an 89 kg male pig, transferred an average of 156 ml/min +/- 39.6 of O2 and 187 ml/min +/- 21.4 of CO2 with an average blood flow rate of 2.5 L/min +/- 0.49, Pn of 0.052 W/(l/min) +/- 0.0067, and R of 3.4 mmHg/(L/min) +/- 0.74. These experiments suggest that such an artificial lung can temporarily support the gas transfer requirements of adult humans without over-loading the right ventricle.

摘要

一种低输入阻抗的胸腔内人工肺正在研发中,用于急性呼吸衰竭或作为移植桥梁。该装置使用微孔中空纤维,纤维束的孔隙率为0.74,表面积为1.83平方米。纤维束置于热成型聚对苯二甲酸乙二醇酯葡萄糖改性外壳内,总体积为800立方厘米。血液进出口为内径18毫米的血管移植物。在进口移植物与装置之间是一个内径1英寸的薄壁乳胶管顺应性腔室。这些装置通过正中胸骨切开术植入约克郡猪体内,与肺动脉和左心房进行端侧吻合。远端肺动脉被阻断,以使右心室输出血流转向该装置。猪1和猪2得到了24小时的完全支持,然后被处死。猪3得到了20小时的部分支持,死于出血并发症。首次植入一只55千克的雄性猪,平均输送氧气176毫升/分钟±42.4,输送二氧化碳190毫升/分钟±39.7,平均血流速度为2.7升/分钟±0.46。归一化平均右心室输出功率Pn为0.062瓦/(升/分钟)±0.0082,平均装置阻力R为3.5毫米汞柱/(升/分钟)±0.62。第二次植入一只60千克的雄性猪,平均输送氧气204毫升/分钟±22.5,输送二氧化碳242毫升/分钟±17.2,平均血流速度为3.7升/分钟±0.45,Pn为0.064瓦/(升/分钟)±0.0067,R为4.3毫米汞柱/(升/分钟)±0.89。第三次植入一只89千克的雄性猪,平均输送氧气156毫升/分钟±39.6,输送二氧化碳187毫升/分钟±21.4,平均血流速度为2.5升/分钟±0.49,Pn为0.052瓦/(升/分钟)±0.0067,R为3.4毫米汞柱/(升/分钟)±0.74。这些实验表明,这种人工肺可以暂时满足成年人类的气体交换需求,而不会使右心室负担过重。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验