Hemodynamic evaluation of a new bileaflet valve prosthesis: an acute animal experimental study.

作者信息

Hasenkam J M, Nygaard H, Terp K, Riis C, Paulsen P K

机构信息

Department of Cardiothoracic and Vascular Surgery, Skejby Sygehus, Aarhus University Hospital, Denmark.

出版信息

J Heart Valve Dis. 1996 Nov;5(6):574-80.

PMID:8953434
Abstract

BACKGROUND AND AIMS OF THE STUDY

A newly developed heart valve (Medtronic Parallel) was tested in an acute animal experimental model.

METHODS

Five prototype valves were implanted in the aortic position in seven 90 kg pigs to enable acute evaluation of the hemodynamic performance in terms of turbulent stresses and transvalvular pressure drop. Turbulent stresses in the ascending aorta were measured using a 10 MHz perivascular Doppler echocardiographic transducer designed to measure the radical velocity component at 17 different points covering the aortic cross-sectional area.

RESULTS

The drop in transvalvular peak pressure measured with fluid-filled catheters showed a non-linear relationship with cardiac output and was always < 12 mmHg. The Reynolds normal stresses were < 60 N/m2 in systole within 50 ms time windows, which is insufficient to cause mechanical damage of the formed elements of the blood.

CONCLUSIONS

From a hemodynamic point of view the performance of the Medtronic Parallel aortic valve is fully acceptable and within the range of other similar, currently available mechanical valves.