[High resolution computerized tomography: dose evaluation with an anthropomorphic phantom].

作者信息

Malgieri F, Guidi G, Orientale A, Scanni M, Cusati B, Rotondo A, Smaltino F

机构信息

Istituto di Scienze Radiologiche, Facoltà di Medicina e Chirurgia, Università degli Studi Federico II, Napoli.

出版信息

Radiol Med. 1996 Jul-Aug;92(1-2):110-3.

PMID:8966248
Abstract

High resolution Computed Tomography (HRCT), a technique for pulmonary parenchyma studies, has been for a long time performed with only few scans with a large interval in between: the reasons were the supposed increase in the dose to the patient after kVp and mAs increase, which is necessary to reduce background noise, and the characteristic distribution of most diffuse lung diseases. Of late, HRCT skin dose has been demonstrated to be markedly lower than that of conventional CT (8/10 mm contiguous slices). To evaluate organ doses to patients during HRCT, we compared conventional CT (10 mm slices), HRCT (1.5 mm slices) and chest X-ray doses, as measured on an anthropomorphous phantom. We tested several scan protocols, with different kVp and mAs values to simulate various clinical situations for the two techniques, always with both slice thickness values and both contiguous and non-contiguous (10 mm gap) slices for HRCT scans. Doses were substantially similar for the two methods with contiguous slices in every protocol. "Clinical" HRCT dose (non-contiguous slices with 10 mm gap) is much lower than that of conventional CT--i.e., about 13% of conventional CT dose. On the other hand, HRCT dose is still two orders of quantity higher than that of standard radiography. To conclude, the importance of HRCT in pulmonary parenchyma studies is confirmed.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验