Eto A, Oishi T, Nakano N I, Chikaura Y
First Production Division, Chemo-Sero-Therapeutic Research Institute (KAKETSUKEN), Kumamoto, Japan.
Clin Chem. 1996 Dec;42(12):2008-14.
Seraclear-HE, containing gamma-glutamyltransferase (GGT, EC 2.3.2.2) derived from the cell culture of a human macrophage cell line, was evaluated as a candidate enzyme reference material (ERM) to calibrate routine methods in terms of a Reference Method for GGT. We have compared this preparation with commercially available materials, including Certified Reference Material 319, at 30 degrees C and 37 degrees C with respect to kinetic properties and with respect to commutability between the Reference Method and each of 15 analytical procedures involving five structurally different donor substrates. GGTs of human origin are far more commutable with reagents of varied types than are GGTs derived from animals. Calibration of 44 patients' sera with Seraclear-HE decreased average intermethod variation from 20% to approximately 4%, whereas GGTs of animal origin showed intermethod variations of approximately 30% with no benefit from calibration. Seraclear-HE is promising as a secondary or working ERM to be used as an intermethod calibrator.
含有源自人巨噬细胞系细胞培养物的γ-谷氨酰转移酶(GGT,EC 2.3.2.2)的Seraclear-HE,作为候选酶参考物质(ERM)进行了评估,以根据GGT的参考方法校准常规方法。我们已将该制剂与市售材料(包括有证参考物质319)在30℃和37℃下就动力学性质以及参考方法与涉及五种结构不同供体底物的15种分析程序中的每一种之间的互换性进行了比较。与源自动物的GGT相比,人源GGT与各种类型试剂的互换性要强得多。用Seraclear-HE校准44例患者血清可使平均方法间差异从20%降至约4%,而源自动物的GGT方法间差异约为30%,校准无益处。Seraclear-HE作为二级或工作ERM用作方法间校准物很有前景。