Prototype instrument employing a microsensor array for the analysis of organic vapors in exhaled breath.

作者信息

Groves W A, Zellers E T

机构信息

Department of Environmental and Industrial Health, School of Public Health, University of Michigan, Ann Arbor 48109-2029, USA.

出版信息

Am Ind Hyg Assoc J. 1996 Dec;57(12):1103-8. doi: 10.1080/15428119691014189.

Abstract

A prototype portable instrument capable of selectively measuring organic vapors in breath at low- and sub-ppm concentrations was tested. The instrument employs an array of four polymer-coated surface-acoustic-wave (SAW) resonators and microprocessor-controlled pumps, valves, and heating elements for preconcentration and thermal desorption of breath samples. Response data are passed to an external personal computer for processing. Sensor responses are based on the reversible changes in the mass and viscoelasticity of the polymeric coatings accompanying vapor sorption. Calibrations were performed for perchloroethylene, trichloroethylene, and methoxyflurane at 100% relative humidity using 15 mg of TenaxTA as the preconcentrator adsorbent. Analyses were then conducted of 250-mL breath samples prepared in Tedlar bags spiked with one of the three solvents within the range of 2 to 18 ppm. Differences between concentrations measured using the prototype instrument and those measured by gas chromatography ranged from -6 to 17% for perchloroethylene. Similar results were obtained for trichloroethylene and methoxyflurane with larger differences occurring at lower concentrations. Estimated limits of detection of 0.7, 0.6, and 4 ppm were achieved for perchloroethylene, trichloroethylene, and methoxyflurane, respectively. These preliminary results demonstrate the feasibility of using SAW sensor technology for field breath analysis.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验