Riedy L W, Walter J S
Rehabilitation Research and Development Center, Hines VA Hospital, IL 60141, USA.
IEEE Trans Biomed Eng. 1996 Jun;43(6):660-3. doi: 10.1109/10.495287.
The safe charge injection density for pulsing of 316LVM electrodes has been reported to be 40 microC/cm2. However, only 20 microC/cm2 is available for nonfaradic charge transfer and double layer charge injection. Therefore, we evaluated long term pulsing at 20 microC/cm2 with capacitor coupling.
据报道,316LVM电极脉冲的安全电荷注入密度为40微库仑/平方厘米。然而,只有20微库仑/平方厘米可用于非法拉第电荷转移和双层电荷注入。因此,我们评估了采用电容耦合在20微库仑/平方厘米下的长期脉冲情况。