Yang Z, Suzuki H, Sasaki S, Karube I
Research Center for Advanced Science and Technology, University of Tokyo, Japan.
Appl Microbiol Biotechnol. 1996 Aug;46(1):10-4. doi: 10.1007/s002530050776.
Disposable-type microbial sensors were prepared for the determination of biochemical oxygen demand (BOD). The yeast, Trichosporon cutaneum, was directly immobilized on the surface of miniature oxygen electrodes using an ultraviolet crosslinking resin (ENT-3400). The oxygen electrodes (15 mm x 2 mm x 0.4 mm) were made on silicon substrates using micromachining techniques. They were Clark-type two-electrode systems with-1021 mV applied to the working electrode. Typical response times of the BOD sensors were in the range of 7-20 min. At 20 degrees C, the sensors' dynamic range was from 0 to 18 mg/1 BOD when a glucose/glutamate BOD standard solution was used. The lower limit of detection was 0.2 mg/l BOD. This value was one order of magnitude lower than that of sensors previously reported. The sensors' operational lifetime of 3 days was satisfactory for a disposable type. The sensors' responses were reproducible to within 8% relative standard deviation. The BOD sensors' were applied to untreated and treated waste waters from industrial effluents and municipal sewage. BOD values determined using these sensors correlated well with those determined by the conventional 5-day BOD determination method.
制备了一次性微生物传感器用于测定生化需氧量(BOD)。使用紫外线交联树脂(ENT - 3400)将皮状丝孢酵母直接固定在微型氧电极表面。氧电极(15毫米×2毫米×0.4毫米)采用微加工技术在硅基片上制成。它们是克拉克型双电极系统,工作电极施加 - 1021毫伏电压。BOD传感器的典型响应时间在7 - 20分钟范围内。在20摄氏度时,当使用葡萄糖/谷氨酸BOD标准溶液时,传感器的动态范围为0至18毫克/升BOD。检测下限为0.2毫克/升BOD。该值比先前报道的传感器低一个数量级。对于一次性类型的传感器,其3天的使用寿命是令人满意的。传感器的响应具有可重复性,相对标准偏差在8%以内。BOD传感器应用于工业废水和城市污水的未处理和处理后的废水。使用这些传感器测定的BOD值与通过传统5天BOD测定方法测定的值相关性良好。