Suppr超能文献

一种用于检测深静脉血栓形成的廉价B型超声设备的验证

Validation of an inexpensive B-mode ultrasound device for detection of deep vein thrombosis.

作者信息

Trottier S J, Todi S, Veremakis C

机构信息

Department of Critical Care Medicine, St. John's Mercy Medical Center/St. Louis University, MO, USA.

出版信息

Chest. 1996 Dec;110(6):1547-50. doi: 10.1378/chest.110.6.1547.

Abstract

OBJECTIVE

To validate a portable, inexpensive, real-time, B-mode ultrasound device compared with duplex ultrasound in the detection of proximal lower extremity deep vein thrombosis in hospitalized patients clinically suspected of having deep vein thrombosis.

DESIGN

Prospective cohort study.

SETTING

Tertiary care community teaching hospital.

PATIENTS

Medical-surgical hospitalized patients undergoing duplex ultrasonography for clinically suspected lower extremity deep vein thrombosis.

INTERVENTIONS

Hospitalized patients who underwent duplex ultrasound examinations were enrolled in the study. Blinded from the duplex ultrasound results, the investigators utilized the study ultrasound device to perform compression ultrasonography of the common femoral, superficial femoral, and popliteal veins within 48 h of the duplex examinations. The results of the study ultrasound device were recorded as normal (compressible) or abnormal (noncompressible).

RESULTS

Of the 198 lower limbs evaluated, duplex ultrasonography documented 34 proximal lower extremity deep vein thrombi. The study ultrasound device detected 32 of the 34 proximal thrombi detected by duplex ultrasonography. One false-positive result of an examination occurred with the study ultrasound device. Compared with duplex ultrasonography, the study ultrasound device had a sensitivity of 94%, specificity of 99%, positive predictive value of 97%, and negative predictive value of 98%.

CONCLUSIONS

The results of this investigation document that the study ultrasound device has an acceptable sensitivity, specificity, and diagnostic accuracy for clinical use in detection of proximal lower extremity deep vein thrombosis. Further evaluation and validation of this ultrasound device are warranted.

摘要

目的

验证一种便携式、低成本、实时的B型超声设备与双功超声在检测临床怀疑患有深静脉血栓形成的住院患者下肢近端深静脉血栓方面的效果。

设计

前瞻性队列研究。

地点

三级医疗社区教学医院。

患者

因临床怀疑下肢深静脉血栓形成而接受双功超声检查的内科和外科住院患者。

干预措施

对接受双功超声检查的住院患者进行研究。研究人员在不知道双功超声检查结果的情况下,使用研究用超声设备在双功检查后48小时内对股总静脉、股浅静脉和腘静脉进行压迫超声检查。将研究用超声设备的检查结果记录为正常(可压缩)或异常(不可压缩)。

结果

在评估的198条下肢中,双功超声检查发现34例下肢近端深静脉血栓。研究用超声设备检测出双功超声检查发现的34例近端血栓中的32例。研究用超声设备出现了1例假阳性检查结果。与双功超声检查相比,研究用超声设备的敏感性为94%,特异性为99%,阳性预测值为97%,阴性预测值为98%。

结论

本研究结果表明,研究用超声设备在检测下肢近端深静脉血栓的临床应用中具有可接受的敏感性、特异性和诊断准确性。有必要对该超声设备进行进一步评估和验证。