[角膜散光对单焦点和多焦点人工晶状体眼对比敏感度的影响——物理眼的理论研究]

[Effect of corneal astigmatism on contrast sensitivity in mono- and multifocal pseudophakia--a theoretical study of the physical eye].

作者信息

Eisenmann D, Wagner R, Dick B, Jacobi K W

机构信息

Universitäts-Augenklinik, Giessen.

出版信息

Klin Monbl Augenheilkd. 1996 Aug-Sep;209(2-3):125-31. doi: 10.1055/s-2008-1035291.

Abstract

BACKGROUND

Several clinical studies indicate, that the visual function of multifocal IOLs (MIOL) is impaired by corneal astigmatism. To assess the influence of uncorrected corneal astigmatism on the contrast sensitivity function (CSF) of mono- and multifocal IOLs, an "optical implantation" of physical eyes with astigmatic corneas and IOLs was performed in younger subjects.

METHODS

The virtual image of physical eyes with a 40 dpt achromate as artificial cornea and the (M)IOL in a water bath was projected on the retina of the observer by means of an exactly adjusted video objective. Silicone lenses with defined astigmatisms (+1; +2; +4; +6 dpt) were put in front of the achromate to produce an artificial corneal astigmatism. We compared results of a standard monofocal IOL (Pharmacia 811B), a multizone progressive MIOL (AMO Array SSM-26NB) and a diffractive MIOL (Pharmacia 811E). CSF through these IOLs in the physical eyes was measured in ten healthy subjects (mean age: 27.4 y.) with the B-VAT II-SG Video Acuity Tester (Mentor O&O), which uses sine wave gratings of five different spatial frequencies (1.5; 3; 6; 12; 20 cpd).

RESULTS

Without astigmatic lenses, all IOLs showed a mean CSF within the age-related norm, but the monofocal IOL yielded significantly better results than both MIOLs at three spatial frequencies (3; 6; 12 cpd). With additional astigmatic lenses of 2 dpt and more, mean CSF of all IOLs was below normal range, but there was no difference in the performance of the three lens styles.

CONCLUSION

CSF of MIOLs seems to be less sensitive to uncorrected corneal astigmatism than CSF of the monofocal IOL. This suggests, that a higher preoperative astigmatism does not severely affect the image quality through a multifocal IOL.

摘要

背景

多项临床研究表明,多焦点人工晶状体(MIOL)的视觉功能会受到角膜散光的损害。为了评估未矫正的角膜散光对单焦点和多焦点人工晶状体对比敏感度功能(CSF)的影响,对较年轻受试者中具有散光角膜和人工晶状体的人眼进行了“光学植入”。

方法

通过精确调整的视频物镜,将以40 dpt消色差透镜作为人工角膜以及(M)IOL置于水浴中的人眼虚拟图像投射到观察者的视网膜上。将具有确定散光度数(+1;+2;+4;+6 dpt)的硅胶镜片放在消色差透镜前面,以产生人工角膜散光。我们比较了标准单焦点人工晶状体(Pharmacia 811B)、多区域渐进式MIOL(AMO Array SSM - 26NB)和衍射型MIOL(Pharmacia 811E)的结果。使用B - VAT II - SG视频视力测试仪(Mentor O&O),该仪器使用五种不同空间频率(1.5;3;6;12;20 cpd)的正弦波光栅,对十名健康受试者(平均年龄:27.4岁)通过这些人工晶状体在人眼中的CSF进行了测量。

结果

在没有散光镜片的情况下,所有人工晶状体的平均CSF均在年龄相关的正常范围内,但在三个空间频率(3;6;12 cpd)下,单焦点人工晶状体的结果明显优于两种MIOL。当使用2 dpt及以上的额外散光镜片时,所有人工晶状体的平均CSF均低于正常范围,但三种镜片类型的性能没有差异。

结论

MIOL的CSF似乎比单焦点人工晶状体的CSF对未矫正的角膜散光不太敏感。这表明,术前较高的散光度数不会严重影响通过多焦点人工晶状体的图像质量。