The value of metal artifact reduction in quantitative bone density determination in the vicinity of metallic implants was studied.
Ten cadaver femora with hip endoprostheses implanted intra vitam were examined in CT with varying slice thicknesses, tube voltage, scan time and magnification factor. The artifact-laden images were then processed using a metal artifact reduction programme (MAR). The metallic implant was removed, allowing the bone slices to be examined again with the same imaging parameters, but without the bothersome metal artifacts. This allowed comparison of bone density values obtained before and after application of MAR with data obtained after implant removal.
A comparison of the bone density measurements between the sample group without implants and the group with indwelling implants prior to MAR showed an unexpected correlation coefficient r = 0.975-0.977. The correlation coefficients after MAR rose to r = 0.987-0.992 in 125 kV and r = 0.989-0.991 in 85 kV modes. The reproducibility of the density measurements for the groups without implants and with implants following MAR showed variation coefficients (VC) of 1.16% and 0.75% respectively. The reproducibility of the manual definition of the region of interest (ROI) in bone showed a VC = 2.19%.
Bone density in the vicinity of metallic implants can be reliably and reproducibly determined using MAR.
研究了金属伪影减少在金属植入物附近定量骨密度测定中的价值。
对10具生前植入髋关节假体的尸体股骨进行CT检查,改变切片厚度、管电压、扫描时间和放大倍数。然后使用金属伪影减少程序(MAR)处理带有伪影的图像。移除金属植入物,使骨切片能够在相同成像参数下再次进行检查,但没有烦人的金属伪影。这使得能够将应用MAR前后获得的骨密度值与植入物移除后获得的数据进行比较。
在MAR之前,无植入物的样本组与有植入物的组之间的骨密度测量比较显示,意外的相关系数r = 0.975 - 0.977。在125 kV模式下,MAR后的相关系数升至r = 0.987 - 0.992,在85 kV模式下为r = 0.989 - 0.991。无植入物组和MAR后有植入物组的密度测量再现性分别显示变异系数(VC)为1.16%和0.75%。骨中感兴趣区域(ROI)手动定义的再现性显示VC = 2.19%。
使用MAR可以可靠且可重复地测定金属植入物附近的骨密度。