使用双门控经颅多普勒超声鉴别栓子与伪像。

Differentiation between emboli and artefacts using dual-gated transcranial Doppler ultrasound.

作者信息

Smith J L, Evans D H, Fan L, Bell P R, Naylor A R

机构信息

Department of Surgery, Faculty of Medicine, Leicester University, UK.

出版信息

Ultrasound Med Biol. 1996;22(8):1031-6. doi: 10.1016/s0301-5629(96)00103-2.

Abstract

It is well documented that transcranial Doppler ultrasound has the ability to detect cerebral emboli. During intraoperative patient monitoring studies, many signals due to artefact (probe motion, patient movement or surgical manipulation) are also detected and can be difficult to distinguish from genuine embolic events. We have constructed a Doppler system that can simultaneously range-gate at two separate depths, in order to test the hypothesis that it should be possible to distinguish between emboli and artefact by comparing the signal from the two separate regions within the vessel. The classification algorithm is based on the principle that emboli propagate with blood motion (whereas artefacts do not) and thus will be detected sequentially at different depths along the insonated cerebral artery. One hundred thirty-eight (presumed) embolic and 170 artefact signals were analysed. The median (interquartile range) gate separation was 10.01 mm (7.41-10.78 mm). The time delay between detection of embolic signals in the two channels was 11.04 ms (6.24-16.41 ms, but was only 0.08 ms (-0.48(-)+0.64 ms) for artefact (p < 0.0001). Dual-gated Doppler ultrasound is a conclusive and independent method that differentiates emboli from artefact. Incorporation of this system for long-term monitoring may eliminate the need for an experienced observer to be present.

摘要

有充分文献记载,经颅多普勒超声能够检测到脑栓塞。在术中患者监测研究中,还会检测到许多由伪像(探头移动、患者移动或手术操作)引起的信号,这些信号可能难以与真正的栓塞事件区分开来。我们构建了一个多普勒系统,该系统可以同时在两个不同深度进行距离选通,以便检验这样一个假设:通过比较血管内两个不同区域的信号,应该能够区分栓塞和伪像。分类算法基于这样一个原理:栓塞随血流运动传播(而伪像则不然),因此会在沿被检测脑动脉的不同深度依次被检测到。对138个(假定的)栓塞信号和170个伪像信号进行了分析。门控分离的中位数(四分位间距)为10.01毫米(7.41 - 10.78毫米)。两个通道中检测到栓塞信号的时间延迟为11.04毫秒(6.24 - 16.41毫秒),而伪像的时间延迟仅为0.08毫秒(-0.48(-) + 0.64毫秒)(p < 0.0001)。双门控多普勒超声是一种能够区分栓塞和伪像的确定性独立方法。将该系统用于长期监测可能无需经验丰富的观察者在场。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验