Suppr超能文献

颅内动脉瘤夹的比较研究:闭合和打开力以及物理耐久性。

A comparative study of intracranial aneurysm clips: closing and opening forces and physical endurance.

作者信息

Ooka K, Shibuya M, Suzuki Y

机构信息

Department of Neurosurgery, Nagoya University School of Medicine, Japan.

出版信息

Neurosurgery. 1997 Feb;40(2):318-23. doi: 10.1097/00006123-199702000-00016.

Abstract

OBJECTIVE

The closing and opening forces of various types of intracranial aneurysm clips were measured, and the physical endurance of the clips was evaluated.

METHODS

Straight, bayonet, angled, fenestrated, and temporary Sugita and Yasargil clips were tested by measuring the forces exerted by the blades as they were opened in wider increments and at various points along their length by a specially designed clip tester. Closing forces after repeated openings, continuous opening, or multiple sterilization procedures also were measured.

RESULTS

The opening forces of Yasargil clips increased linearly and more sharply as the blades were opened wider than those of Sugita clips. Only Sugita clips showed significant differences between the opening and closing forces (hysteresis). Closing forces of all of the tested clips were according to a hyperbolic curve. Straight clips showed a sharper rise in closing forces than those of bayonet, angled, or fenestrated clips. A relatively weak closing force at the tip of Yasargil long clips and a strong closing force at the base of temporary clips were unexpected findings. The closing force decreased by only 3 to 7% after 500 openings and did not decrease after continuous opening for 17 hours or 100 sterilization procedures.

CONCLUSION

Our study showed that the closing forces differed depending not only on the distance of the point of measurement from the clip spring but also on the shapes of the clip blades. Repeated opening or sterilization of the clips did not significantly decrease the closing forces of the aneurysm clips.

摘要

目的

测量各种类型颅内动脉瘤夹的开合力,并评估夹子的物理耐久性。

方法

通过一个专门设计的夹子测试仪,测量直形、刺刀形、角形、带孔以及临时使用的杉田(Sugita)和亚萨吉尔(Yasargil)夹子在以更大增量打开时以及沿其长度的不同点处刀片施加的力,对这些夹子进行测试。还测量了在重复打开、持续打开或多次灭菌程序后的闭合力。

结果

随着刀片打开幅度增大,亚萨吉尔夹子的打开力比杉田夹子线性增加得更急剧。只有杉田夹子在打开力和闭合力之间显示出显著差异(滞后现象)。所有测试夹子的闭合力均符合双曲线。直形夹子的闭合力上升比刺刀形、角形或带孔夹子更急剧。亚萨吉尔长夹子尖端闭合力相对较弱以及临时夹子基部闭合力较强是意外发现。在500次打开后闭合力仅下降3%至7%,在持续打开17小时或进行100次灭菌程序后闭合力并未下降。

结论

我们的研究表明,闭合力不仅取决于测量点距夹子弹簧的距离,还取决于夹子刀片的形状。夹子的重复打开或灭菌并未显著降低动脉瘤夹的闭合力。