Yamakoshi Y, Otaki H, Shinozuka N, Masuda H
Faculty of Engineering, Gunma University, Japan.
Ultrasonics. 1996 Dec;34(8):769-75. doi: 10.1016/s0041-624x(96)00080-7.
We have developed a fetal movement monitoring system based on small displacement measurement of internal tissues. When ultrasonic pulses are transmitted to the fetus, the reflected ultrasonic waves which have a Doppler frequency shift due to the fetal movements are detected by using an ultrasonic pulsed Doppler technique. In this paper, we propose a displacement measurement method for internal tissues which is based on the Doppler signal digital detection technique. In the method, the received ultrasonic RF signals are sampled with a sampling frequency of four times higher than the centre frequency of the ultrasonic waves; the Doppler frequency shift signals are derived using digital signal processing. From the detected signals, the internal displacements are estimated using the arc-tangent method. The basic algorithm of the detection method has already been used in the area of blood flow sensing, however, we apply the algorithm to the displacement measurement of internal tissues. The comparison between the proposed method and the conventional method is presented. The fetal movement quantitative monitoring system based on the method which has been constructed is shown.