[胫骨干密度测量检查的可重复性]

[Reproducibility of densitometric examinations of the tibial shaft].

作者信息

Kaleta M, Tokarowski A, Kusz D, Wójcik B, Bozek M

机构信息

Katedry i Kliniki Ortopedii Slaskiej Akademii Medycznej w Katowicach.

出版信息

Chir Narzadow Ruchu Ortop Pol. 1996;61(5):473-8.

PMID:9026419
Abstract

The aim of this paper is to evaluate the reproducibility of tibial shaft densitometry in own method under development. It will be expressed as percentage value of coefficient of variability (%CV). Bone mineral content (BMC) and bone mineral density (BMD) were measured with a dual energy X-ray densitometer Lunar DPX; the data were analyzed with a "Orthopedic" software. Three tibial bone scans were done in 10 males. High reproducibility was found, ranging from 0.98% to 1.45% and from 1.45% to 2.08% for BMC and BMD respectively. Reproducibility increases as the analyzed surface increases. Reproducibility so high allows for further investigations; it might be interesting to monitor BMC and BMD of the tibia in course of fracture treatment by Polfix or Zespol method.

摘要

本文旨在评估正在研发的自身方法中胫骨干密度测定的可重复性。它将以变异系数百分比值(%CV)表示。使用双能X线骨密度仪Lunar DPX测量骨矿物质含量(BMC)和骨密度(BMD);数据用“骨科”软件进行分析。对10名男性进行了三次胫骨骨扫描。发现可重复性很高,BMC和BMD的可重复性分别为0.98%至1.45%和1.45%至2.08%。随着分析表面的增加,可重复性提高。如此高的可重复性允许进行进一步研究;监测通过Polfix或Zespol方法进行骨折治疗过程中胫骨的BMC和BMD可能会很有意思。