Tanaka S, Yamakoshi K
Faculty of Engineering, Yamagata University, Yonezawa, Japan.
Med Biol Eng Comput. 1996 Nov;34(6):441-7. doi: 10.1007/BF02523848.
A portable instrument, based on a volume-compensation technique, is designed for ambulatory monitoring of indirect beat-to-beat blood pressure (BP) in the superficial temporal artery. The instrument consists of a small disc-type cuff and a portable unit carried by the subject. Several components are integrated in the cuff for applying counter-pressure to the artery, i.e. a reflectance-type photo-plethysmographic sensor for arterial volume detection, a pressure sensor for cuff pressure Pc measurement and a nozzle flapper-type- electro-pneumatic convertor for controlling Pc. The portable unit includes volume servo control circuitry and a microprocessor-based signal-processing and recording unit. This automatically performs all the necessary measurement procedures and stores into a memory IC element the processed systolic, mean and diastolic blood pressure data, together with pulse intervals on a beat-to-beat basis from the servo-controlled Pc (indirectly measured BP waveform). With this instrument, momentary changes in BP during ambulatory situations such as bicycle ergometer exercise and daily activities including motorway driving are successfully recorded. From the results of simultaneous measurement of the subject's posture changes, the effect of posture change on blood pressure, e.g. baroreceptor-cardiac reflex, is also clearly demonstrated.
一种基于容积补偿技术的便携式仪器,设计用于动态监测颞浅动脉的逐搏间接血压(BP)。该仪器由一个小型圆盘式袖带和受试者携带的便携式单元组成。袖带中集成了几个部件,用于向动脉施加反压,即用于检测动脉容积的反射型光电容积脉搏波传感器、用于测量袖带压力Pc的压力传感器以及用于控制Pc的喷嘴挡板型电-气转换器。便携式单元包括容积伺服控制电路和基于微处理器的信号处理与记录单元。它自动执行所有必要的测量程序,并将处理后的收缩压、平均压和舒张压数据以及来自伺服控制的Pc(间接测量的BP波形)的逐搏脉搏间期存储到一个存储集成电路元件中。使用该仪器,成功记录了诸如自行车测力计运动和包括高速公路驾驶在内的日常活动等动态情况下血压的瞬间变化。通过同时测量受试者姿势变化的结果,姿势变化对血压的影响,例如压力感受器-心脏反射,也得到了清晰证明。