用于评估基于计算机的摆位误差检测方法的基准测试案例:具有已知摆位误差的逼真数字重建电子射野影像。

Benchmark test cases for evaluation of computer-based methods for detection of setup errors: realistic digitally reconstructed electronic portal images with known setup errors.

作者信息

Fritsch D S, Raghavan S, Boxwala A, Earnhart J, Tracton G, Cullip T, Chaney E L

机构信息

Department of Radiation Oncology, University of North Carolina, Chapel Hill 27599-7512, USA.

出版信息

Int J Radiat Oncol Biol Phys. 1997 Jan 1;37(1):199-204. doi: 10.1016/s0360-3016(96)00479-8.

Abstract

PURPOSE

The purpose of this investigation was to develop methods and software for computing realistic digitally reconstructed electronic portal images with known setup errors for use as benchmark test cases for evaluation and intercomparison of computer-based methods for image matching and detecting setup errors in electronic portal images.

METHODS AND MATERIALS

An existing software tool for computing digitally reconstructed radiographs was modified to compute simulated megavoltage images. An interface was added to allow the user to specify which setup parameter(s) will contain computer-induced random and systematic errors in a reference beam created during virtual simulation. Other software features include options for adding random and structured noise, Gaussian blurring to simulate geometric unsharpness, histogram matching with a "typical" electronic portal image, specifying individual preferences for the appearance of the "gold standard" image, and specifying the number of images generated. The visible male computed tomography data set from the National Library of Medicine was used as the planning image.

RESULTS

Digitally reconstructed electronic portal images with known setup errors have been generated and used to evaluate our methods for automatic image matching and error detection. Any number of different sets of test cases can be generated to investigate setup errors involving selected setup parameters and anatomic volumes. This approach has proved to be invaluable for determination of error detection sensitivity under ideal (rigid body) conditions and for guiding further development of image matching and error detection methods. Example images have been successfully exported for similar use at other sites.

CONCLUSIONS

Because absolute truth is known, digitally reconstructed electronic portal images with known setup errors are well suited for evaluation of computer-aided image matching and error detection methods. High-quality planning images, such as the visible human CT scans from the National Library of Medicine, are essential for producing realistic images. Sets of test cases with systematic and random errors in selected setup parameters and anatomic volumes are suitable for use as standard benchmarks by the radiotherapy community. In addition to serving as an aid to research and development, benchmark images may also be useful for evaluation of commercial systems and as part of a quality assurance program for clinical systems. Test cases and software are available upon request.

摘要

目的

本研究的目的是开发方法和软件,用于计算具有已知设置误差的逼真数字重建电子射野图像,作为基准测试案例,以评估和相互比较基于计算机的电子射野图像匹配和检测设置误差的方法。

方法和材料

对现有的用于计算数字重建射线照片的软件工具进行修改,以计算模拟兆伏级图像。添加了一个界面,允许用户指定在虚拟模拟期间创建的参考射束中哪些设置参数将包含计算机诱导的随机和系统误差。其他软件功能包括添加随机和结构化噪声、高斯模糊以模拟几何不清晰度、与“典型”电子射野图像进行直方图匹配、指定“金标准”图像外观的个人偏好以及指定生成的图像数量。来自美国国立医学图书馆的可见男性计算机断层扫描数据集用作计划图像。

结果

已生成具有已知设置误差的数字重建电子射野图像,并用于评估我们的自动图像匹配和误差检测方法。可以生成任意数量的不同测试案例集,以研究涉及选定设置参数和解剖体积的设置误差。这种方法已被证明对于确定理想(刚体)条件下的误差检测灵敏度以及指导图像匹配和误差检测方法的进一步开发非常有价值。示例图像已成功导出,可供其他机构类似使用。

结论

由于已知绝对真值,具有已知设置误差的数字重建电子射野图像非常适合评估计算机辅助图像匹配和误差检测方法。高质量的计划图像,如来自美国国立医学图书馆的可见人体CT扫描,对于生成逼真图像至关重要。在选定的设置参数和解剖体积中具有系统和随机误差的测试案例集适合放疗界用作标准基准。除了有助于研发外,基准图像还可用于评估商业系统,并作为临床系统质量保证计划的一部分。可根据要求提供测试案例和软件。