Crotti S, Tubiolo D, Pelosi P, Chiumello D, Mascheroni D, Gattinoni L
Istituto di Anestesia e Rianimazione, Università di Milano, Ospedale Maggiore Policlinico IRCCS, Italy.
Int J Artif Organs. 1997 Jan;20(1):22-8.
We compared the performance of a heparinized hollow fiber artificial lung (Medtronic, Minimax) featuring standard hollow fibers (Group A) and experimental hollow fibers with a smaller pore size (Group B). Four sheep in each group underwent a veno-venous bypass for 72 hours. Every 6 hours, at 3 different blood flow rates (BFr) (400, 800, 1200 ml/min), at a constant gas flow rate (Gfr = 4 L/min), and at a constant blood inlet PCO2 (45-55 mmHg), we measured the oxygenation performance (O2 transfer = VO2 and blood outlet PO2 = PO2out), CO2 removal (CO2 transfer = VCO2 and PCO2 outlet = PCO2out) and pressure drop across the device (delta P). A total of about 50 measurement sets were obtained for each group at different time points and blood flow rates. Both groups showed a good oxygenation performance (PO2out always higher than 200 mmHg) and no differences were observed between the two groups (at 1200 ml/min BFr, the average VO2 of all time points was 47 +/- 15 ml/min in group A and 44 +/- 11 ml/min in group B, mean +/- SD, NS). During the first 24 hours, the VCO2 was higher in Group B than in Group A at each BFr (at 1200 ml/min BFr, 81 +/- 18 vs 67 +/- 20 ml/min, p < 0.01), while no differences were observed during the subsequent 48 hours. Throughout the entire experiment, VCO2 increased with increasing BFr in both groups, (in group B, from 43 +/- 14 ml/min at 400 ml/min BFr, to 73 +/- 17 ml/min at 1200 ml/min BFr, average of all time points, p < 0.01). In both groups the delta P increased with the increasing BFr, but it was lower in Group B than in Group A at BFr 800 and 1200 ml/min (at 1200 ml/min BFr, 51 +/- 15 mmHg vs 65 +/- 17 mmHg, p < 0.01), and remained stable for the entire experimental period.
我们比较了配备标准中空纤维的肝素化中空纤维人工肺(美敦力公司,Minimax)(A组)和具有较小孔径的实验性中空纤维人工肺(B组)的性能。每组4只绵羊接受静脉-静脉旁路循环72小时。每6小时,在3种不同的血流速率(BFr)(400、800、1200毫升/分钟)、恒定的气体流速(Gfr = 4升/分钟)以及恒定的血液入口PCO2(45 - 55毫米汞柱)条件下,我们测量了氧合性能(氧输送 = VO2,血液出口PO2 = PO2out)、二氧化碳清除(二氧化碳输送 = VCO2,PCO2出口 = PCO2out)以及设备两端的压降(ΔP)。在不同时间点和血流速率下,每组共获得约50组测量数据。两组均表现出良好的氧合性能(PO2out始终高于200毫米汞柱),两组之间未观察到差异(在1200毫升/分钟BFr时,A组所有时间点的平均VO2为47 ± 15毫升/分钟,B组为44 ± 11毫升/分钟,均值 ± 标准差,无显著差异)。在最初的24小时内,每个BFr下B组的VCO2均高于A组(在1200毫升/分钟BFr时,分别为81 ± 18与67 ± 20毫升/分钟,p < 0.01),而在随后的48小时内未观察到差异。在整个实验过程中,两组的VCO2均随BFr的增加而升高(在B组中,从400毫升/分钟BFr时的43 ± 14毫升/分钟,升至1200毫升/分钟BFr时的73 ± 17毫升/分钟,所有时间点的平均值,p < 0.01)。两组的ΔP均随BFr的增加而升高,但在800和1200毫升/分钟BFr时,B组的ΔP低于A组(在1200毫升/分钟BFr时,分别为51 ± 15毫米汞柱与65 ± 17毫米汞柱,p < 0.01),并且在整个实验期间保持稳定。