Villafana T
Acta Radiol Ther Phys Biol. 1977 Jun;16(3):281-8. doi: 10.3109/02841867709133948.
The modulation transfer function of radiologic imaging systems is commonly obtained by determining the line spread function (LSF) of the system and computing its Fourier transform. Ideally, LSF should be obtained with infinitesimally narrow slits. The use of finite slits for obtaining the LSF is analyzed theoretically. An expression for the MTF of a finite rectangular slit is derived. Slit width correction of observed MTF is discussed as well as the correction for any generalized slit configuration.